Characterizing a visual lateral flow device for rapid SARS-CoV-2 virus protein detection: pre-clinical and system assessment
Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) infections have affected more than 769 million individuals worldwide over the last few years. Although the pandemic is transitioning into an endemic, the COVID-19 outbreak is still a global concern. A rapid screening platform is needed for...
Gespeichert in:
Veröffentlicht in: | Analytical methods 2024-05, Vol.16 (17), p.274-275 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) infections have affected more than 769 million individuals worldwide over the last few years. Although the pandemic is transitioning into an endemic, the COVID-19 outbreak is still a global concern. A rapid screening platform is needed for effective preventive and control measures. Herein, a visual rapid lateral flow platform for SARS-CoV-2 nucleocapsid protein detection is developed. Under optimal conditions, the system demonstrated good detection sensitivity and selectivity against tested respiratory viruses. The system provides direct visual detection with a limit of 0.7 ng of the nucleocapsid protein per mL of a sample (0.7 ng mL
−1
) within 15 minutes. Further, a correlation between direct visual detection and semi-quantitative analysis using a reader showed a similar detection limit (
R
2
= 0.9571). The repeatability and reproducibility studies highlighted the potential of the system for the rapid screening of SARS-CoV-2 infection, with variations within 5% and 10% at high and low protein concentrations, respectively. Subsequent pre-clinical validation to correlate the performance with the standard molecular approach (RT-PCR) using 170 nasopharyngeal swabs demonstrated 98% estimated sensitivity (95% CI, 89.35-99.95%) and 100% specificity (95% CI, 96.38-100%). The positive and negative predictive values were reported to be 100% and 99%, respectively, with an accuracy of 99.3%. With high viral load samples (Ct value ≤25,
n
= 47), the system demonstrated 100% detection sensitivity and specificity. The proposed technique provides a valuable platform for potential use in rapid screening, particularly during pandemics, where diagnostic capacity and mass screening are crucial.
Analytical and pre-clinical evaluation of a visual screening lateral flow device for SARS-CoV-2 virus protein detection and its system characterization. |
---|---|
ISSN: | 1759-9660 1759-9679 |
DOI: | 10.1039/d3ay02075d |