Determination of aluminum in moisturizing body lotions using graphite furnace atomic absorption spectrometry
A method has been developed for the determination of Al in commercial moisturizing body lotions using high-resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GF AAS). The most sensitive absorption line at 309.271 nm has been used for all determinations. The samples we...
Gespeichert in:
Veröffentlicht in: | Analytical methods 2015-01, Vol.7 (22), p.9636-964 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A method has been developed for the determination of Al in commercial moisturizing body lotions using high-resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GF AAS). The most sensitive absorption line at 309.271 nm has been used for all determinations. The samples were prepared
via
a suspension with 0.3 mol L
−1
nitric acid. Comparisons were made with pyrolysis and atomization curves without a modifier and using ruthenium and zirconium as permanent modifiers, to reduce the interaction of Al with the graphite surface and to improve the atomization efficiency of Al. 400 μg Zr was finally chosen as the best permanent modifier. The limits of detection and quantification were 30 ng g
−1
and 95 ng g
−1
Al, respectively. The proposed method was applied for 15 commercial moisturizing body lotions with results ranging between 0.17 ± 0.05 μg g
−1
and 11.8 ± 0.43 μg g
−1
. Analyte addition tests have been used to evaluate the accuracy of the method and the recovery ranged between 103 and 110%, indicating that there was no serious matrix effect.
Aluminum has been determined in body lotions after mixing with dilute nitric acid using graphite furnace AAS and Zr as a permanent modifier. |
---|---|
ISSN: | 1759-9660 1759-9679 |
DOI: | 10.1039/c5ay02123e |