Re-assessing melt versus solution 13C NMR spectral sensitivity on polyethyleneElectronic supplementary information (ESI) available: Molecular characteristics of PE sample material Marlex® resin EHM 6007; guidelines for selecting baseline regions for a given peak; OSW calculations; NOE considerations; overall u (δ+δ+) profiles in absolute intensity mode; illustration of IW effect on σ (δ+δ+); Detection Limit (DL) comparisons between solution and melt for CH, αδ+, and βδ+ peaks. See DOI: 10.1039/c

At run times of 3 hours to 10 days, with continuous dipolar decoupling, spectral sensitivities at three solution concentrations (with a 10 mm probe) were compared to that of a melt sample (with a 7 mm rotor) of the same polyethylene material. Contrary to all previous findings, solution 13 C NMR was...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Hou, Liping, Guo, Meifang, Qiao, Jinliang, Hsieh, Eric T
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:At run times of 3 hours to 10 days, with continuous dipolar decoupling, spectral sensitivities at three solution concentrations (with a 10 mm probe) were compared to that of a melt sample (with a 7 mm rotor) of the same polyethylene material. Contrary to all previous findings, solution 13 C NMR was found to be superior to the melt technique at all run times. A common mistake in signal-to-noise ratio (SNR) assessment, mixing integral and line height in the same formulation, was corrected. The Observable Sample Weight (OSW) and the extent of nuclear Overhauser enhancement actually achieved were identified as the predominant factors that affected the "signal strength". "Noise level", measured using the revised SNR method, was found to be closely related to the width of the peak integral, which increased with increasing solution/melt (liquid) viscosity. The melt technique did not offer any gains in OSW, but there were significant losses in spectral resolution even with the benefits of magic angle spinning and high power dipolar decoupling. All this means that the Solution technique is still the superior 13 C NMR technique for quantifying molecular structures of soluble polyethylene (PE) materials. At run times of 3 hours to 10 days, with continuous dipolar decoupling, spectral sensitivities at three solution concentrations (with a 10 mm probe) were compared to that of a melt sample (with a 7 mm rotor) of the same polyethylene material.
ISSN:1759-9954
1759-9962
DOI:10.1039/c2py20119d