Closed-Loop Oxygen Control Using a Novel Nasal High-Flow Device: A Randomized Crossover Trial
Oxygen administration is recommended for patients with hypoxemia to achieve a target [Formula: see text] range. Strategies to achieve this in clinical practice are suboptimal. We investigated automatic oxygen titration using a novel nasal high-flow device with closed-loop oxygen control. The objecti...
Gespeichert in:
Veröffentlicht in: | Respiratory care 2021-03, Vol.66 (3), p.416-424 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Oxygen administration is recommended for patients with hypoxemia to achieve a target [Formula: see text] range. Strategies to achieve this in clinical practice are suboptimal. We investigated automatic oxygen titration using a novel nasal high-flow device with closed-loop oxygen control. The objective of this proof-of-concept study was to determine whether closed-loop control was able to respond to desaturation and subsequent recovery in a controlled laboratory-based environment.
We conducted a single-blind randomized crossover trial in adults with chronic respiratory disease who had a resting [Formula: see text] ≥ 92% and desaturated to < 90% during a 6-min walk test (6MWT). Nasal high-flow was administered during a 6MWT and a subsequent 10-min rest period with either room air, a fixed concentration of 28% oxygen, or oxygen titrated automatically using closed-loop control.
The study involved 42 subjects. Closed-loop control maintained [Formula: see text] within the target range of 92-96% for a mean (SD) duration of 54.4 ± 30.1% of the 6MWT and 67.3 ± 26.8% of the recovery period. The proportion of time spent with an [Formula: see text] in the target range during the 6MWT was significantly greater for closed-loop control compared to room air, with a difference of 26.0% (95% CI 17.7-34.2,
< .001); this proportion of time was not significantly different compared to the fixed concentration of 28% oxygen, with a difference of -8.2% (95% CI -16.5 to 0.1,
= .052). The proportion of time spent in the target range during the rest period was significantly greater compared to 28% oxygen, with a difference of 19.3% (95% CI 8.9-29.7,
< .001); this proportion of time was not significantly different compared to room air, with a difference of -9.3% (95% CI -19.7 to 1.0,
= .08).
This study provides proof-of-concept evidence that the novel nasal high-flow device with closed-loop control can respond to changes in [Formula: see text] outside a target saturation range using a model of exercise-induced desaturation and subsequent recovery. |
---|---|