Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems
Purpose To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices. Patients and methods A prospective randomized cohort study, conducted on 47 eyes of 47 healthy orthotropic individuals, with an age range of 18–50 years and a corrected Snellen’s dista...
Gespeichert in:
Veröffentlicht in: | International ophthalmology 2023-02, Vol.43 (2), p.381-386 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices.
Patients and methods
A prospective randomized cohort study, conducted on 47 eyes of 47 healthy orthotropic individuals, with an age range of 18–50 years and a corrected Snellen’s distance visual acuity (CDVA) of 0.8 decimal or better. Angle Kappa is assessed directly using Orbscan® III software version 1.8.165.1. (Bausch and Lomb Rochester, New York, United States), while Pentacam® HR 1.21r.65 (Oculus Optikgeräte GmbH, Wetzlar, Germany) and Sirius device (CSO, version 3.2.1.60, Costruzione Strumenti Oftalmici, Florence, Italy) were used to calculate angle kappa indirectly.
Results
Least mean difference of estimated angle Kappa was between Orbscan and Pentacam devices (− 0.18° ± 1.8), and it was statistically insignificant (
p
value = 0.1294). Differences between both Orbscan and Sirius, and Pentacam and Sirius were statistically significant (
p
value = 0.0004 and |
---|---|