Minifragment plating of the fibula in unstable ankle fractures
Introduction Only 6.4–17% of the load is transmitted through the fibula when weight-bearing. Plate fixation of distal fibular fractures using minifragments (≤ 2.8 mm) could lead to similar reduction with less implant removal (IR) rates, compared to small-fragment plates (3.5 mm). We hypothesized tha...
Gespeichert in:
Veröffentlicht in: | Archives of orthopaedic and trauma surgery 2023-03, Vol.143 (3), p.1499-1504 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
Only 6.4–17% of the load is transmitted through the fibula when weight-bearing. Plate fixation of distal fibular fractures using minifragments (≤ 2.8 mm) could lead to similar reduction with less implant removal (IR) rates, compared to small-fragment plates (3.5 mm). We hypothesized that the use of minifragment plates is at least similar in unscheduled secondary surgery.
Materials and methods
In this retrospective cohort study, all patients with surgically treated distal fibular fractures between October 2015 and March 2021 were included. Patients treated with plate fixation using minifragments and patients treated with small-fragment plates were compared regarding the following outcomes: secondary dislocation, malreduction, implant malposition, nonunion, surgical site infections (SSI) and IR.
Results
Sixty-five patients (54.2%) received a minifragment implant (≤ 2.8 mm) and 55 patients (45.8%) received a small-fragment implant (3.5 mm). There were no patients needing secondary surgery in the minifragment group compared to 9 patients following fixation using small-fragment implants (3 with secondary dislocation, 5 with malreduction and 1 with malposition,
p
= 0.001). SSI rates were 3.1% for minifragment and 9.1% for small-fragment implants (
p
= 0.161). Implant removal was performed significantly less often following use of minifragment implants (17.8% and 53.2%,
p
|
---|---|
ISSN: | 1434-3916 0936-8051 1434-3916 |
DOI: | 10.1007/s00402-022-04397-2 |