Development of a three‐dimensional dose evaluation method for computed tomography
During a single scan using computed tomography, an X‐ray tube orbits along a 360°‐circular path around the patient. A scan obtained using the half‐cylindrical type phantoms with a radiochromic film sandwiched in between reveals a pixel value map illustrating the two‐dimensional (2D) dose distributio...
Gespeichert in:
Veröffentlicht in: | Journal of Applied Clinical Medical Physics 2023-02, Vol.24 (2), p.e13897-n/a |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | During a single scan using computed tomography, an X‐ray tube orbits along a 360°‐circular path around the patient. A scan obtained using the half‐cylindrical type phantoms with a radiochromic film sandwiched in between reveals a pixel value map illustrating the two‐dimensional (2D) dose distribution. A three‐dimensional (3D) dose distribution can be obtained with a 360° rotation of the 2D dose map. This study evaluates the concept and methodology of creating a 3D dose map to develop a phantom with a radiochromic film for obtaining the 3D dose distribution. The coronal and axial plane dose distributions were also evaluated. A single scan computed tomography image obtained using a half‐cylindrical type of acrylic phantom with a sandwiched radiochromic film was studied. The diameters of the phantoms were 10 and 16 cm, and their lengths were 30 cm. A 2D image of the XR‐QA2 film was obtained using an image scanner and image processing software. A red channel image was used to obtain the 3D dose distribution using a computing platform. A pseudo color was applied to the red channel image from which cross‐sectional color images were obtained. Half of the cross‐sectional pixel data were rotated by 360° to generate the data for each axial plane. The image created was saved, and a 3D pixel value map was constructed. The dose measurement procedure for the 3D dose distribution was developed using half‐cylindrical acrylic phantoms with a radiochromic film. |
---|---|
ISSN: | 1526-9914 1526-9914 |
DOI: | 10.1002/acm2.13897 |