Visual observation of extraction of a Micra leadless pacemaker from a human cadaver

Introduction In this article we present the extraction of a Micra from a human cadaver implanted 3 years previously with both visual and X‐ray imaging taken during the removal. Methods A Micra pacemaker was extracted from a human cadaver with endoscopy and fluoroscopy using a Micra delivery tool. Hi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2022-09, Vol.45 (9), p.1056-1061
Hauptverfasser: Bonner, Matthew, Hilpisch, Kathryn, Wika, Kent, Mesich, Mary Lauren, Harris, Megan, White, Troy
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Introduction In this article we present the extraction of a Micra from a human cadaver implanted 3 years previously with both visual and X‐ray imaging taken during the removal. Methods A Micra pacemaker was extracted from a human cadaver with endoscopy and fluoroscopy using a Micra delivery tool. Histological analysis was performed on slices from the tissue surrounding the Micra. Results The fully encapsulated Micra was easily retrieved with a maximum force of 1.9 pounds. Conclusions Even though the Micra was implanted almost 3 years previously, the snaring and extraction of the Micra was performed relatively easily and with minimal force required.
ISSN:0147-8389
1540-8159
DOI:10.1111/pace.14559