Introduction In this article we present the extraction of a Micra from a human cadaver implanted 3 years previously with both visual and X‐ray imaging taken during the removal. Methods A Micra pacemaker was extracted from a human cadaver with endoscopy and fluoroscopy using a Micra delivery tool. Hi...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 2022-09, Vol.45 (9), p.1056-1061 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
In this article we present the extraction of a Micra from a human cadaver implanted 3 years previously with both visual and X‐ray imaging taken during the removal.
Methods
A Micra pacemaker was extracted from a human cadaver with endoscopy and fluoroscopy using a Micra delivery tool. Histological analysis was performed on slices from the tissue surrounding the Micra.
Results
The fully encapsulated Micra was easily retrieved with a maximum force of 1.9 pounds.
Conclusions
Even though the Micra was implanted almost 3 years previously, the snaring and extraction of the Micra was performed relatively easily and with minimal force required. |
---|---|