Analogue Mean Systemic Filling Pressure: a New Volume Management Approach During Percutaneous Left Ventricular Assist Device Therapy
The absence of an accepted gold standard to estimate volume status is an obstacle for optimal management of left ventricular assist devices (LVADs). The applicability of the analogue mean systemic filling pressure (Pmsa) as a surrogate of the mean circulatory pressure to estimate volume status for p...
Gespeichert in:
Veröffentlicht in: | Journal of cardiovascular translational research 2022-12, Vol.15 (6), p.1455-1463 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | The absence of an accepted gold standard to estimate volume status is an obstacle for optimal management of left ventricular assist devices (LVADs). The applicability of the analogue mean systemic filling pressure (Pmsa) as a surrogate of the mean circulatory pressure to estimate volume status for patients with LVADs has not been investigated. Variability of flows generated by the Impella CP, a temporary LVAD, should have no physiological impact on fluid status. This translational interventional ovine study demonstrated that Pmsa did not change with variable circulatory flows induced by a continuous flow LVAD (the average dynamic increase in Pmsa of 0.20 ± 0.95 mmHg from zero to maximal Impella flow was not significant (
p
= 0.68)), confirming applicability of the human Pmsa equation for an ovine LVAD model. The study opens new directions for future translational and human investigations of fluid management using Pmsa for patients with temporary LVADs. |
---|---|