A smartphone-interfaced, low-cost colorimetry biosensor for selective detection of bronchiectasis via an artificial neural network
Exhaled breath (EB) contains several macromolecules that can be exploited as biomarkers to provide clinical information about various diseases. Hydrogen peroxide (H 2 O 2 ) is a biomarker because it indicates bronchiectasis in humans. This paper presents a non-invasive, low-cost, and portable quanti...
Gespeichert in:
Veröffentlicht in: | RSC advances 2022-08, Vol.12 (37), p.23946-23955 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Exhaled breath (EB) contains several macromolecules that can be exploited as biomarkers to provide clinical information about various diseases. Hydrogen peroxide (H
2
O
2
) is a biomarker because it indicates bronchiectasis in humans. This paper presents a non-invasive, low-cost, and portable quantitative analysis for monitoring and quantifying H
2
O
2
in EB. The sensing unit works on colorimetry by the synergetic effect of eosin blue, potassium permanganate, and starch-iodine (EPS) systems. Various sampling conditions like pH, response time, concentration, temperature and selectivity were examined. The UV-vis absorption study of the assay showed that the dye system could detect as low as ∼0.011 ppm levels of H
2
O
2
. A smart device-assisted detection unit that rapidly detects red, green and blue (RGB) values has been interfaced for practical and real-time application. The RGB value-based quantification of the H
2
O
2
level was calibrated against NMR spectroscopy and exhibited a close correlation. Further, we adopted a machine learning approach to predict H
2
O
2
concentration. For the evaluation, an artificial neural network (ANN) regression model returned 0.941
R
2
suggesting its great prospect for discrete level quantification of H
2
O
2
. The outcomes exemplified that the sensor could be used to detect bronchiectasis from exhaled breath. |
---|---|
ISSN: | 2046-2069 2046-2069 |
DOI: | 10.1039/d2ra03769f |