Performance of a Chest Radiograph AI Diagnostic Tool for COVID-19: A Prospective Observational Study

To conduct a prospective observational study across 12 U.S. hospitals to evaluate real-time performance of an interpretable artificial intelligence (AI) model to detect COVID-19 on chest radiographs. A total of 95 363 chest radiographs were included in model training, external validation, and real-t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Radiology. Artificial intelligence 2022-07, Vol.4 (4), p.e210217
Hauptverfasser: Sun, Ju, Peng, Le, Li, Taihui, Adila, Dyah, Zaiman, Zach, Melton-Meaux, Genevieve B, Ingraham, Nicholas E, Murray, Eric, Boley, Daniel, Switzer, Sean, Burns, John L, Huang, Kun, Allen, Tadashi, Steenburg, Scott D, Gichoya, Judy Wawira, Kummerfeld, Erich, Tignanelli, Christopher J
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To conduct a prospective observational study across 12 U.S. hospitals to evaluate real-time performance of an interpretable artificial intelligence (AI) model to detect COVID-19 on chest radiographs. A total of 95 363 chest radiographs were included in model training, external validation, and real-time validation. The model was deployed as a clinical decision support system, and performance was prospectively evaluated. There were 5335 total real-time predictions and a COVID-19 prevalence of 4.8% (258 of 5335). Model performance was assessed with use of receiver operating characteristic analysis, precision-recall curves, and F1 score. Logistic regression was used to evaluate the association of race and sex with AI model diagnostic accuracy. To compare model accuracy with the performance of board-certified radiologists, a third dataset of 1638 images was read independently by two radiologists. Participants positive for COVID-19 had higher COVID-19 diagnostic scores than participants negative for COVID-19 (median, 0.1 [IQR, 0.0-0.8] vs 0.0 [IQR, 0.0-0.1], respectively; < .001). Real-time model performance was unchanged over 19 weeks of implementation (area under the receiver operating characteristic curve, 0.70; 95% CI: 0.66, 0.73). Model sensitivity was higher in men than women ( = .01), whereas model specificity was higher in women ( = .001). Sensitivity was higher for Asian ( = .002) and Black ( = .046) participants compared with White participants. The COVID-19 AI diagnostic system had worse accuracy (63.5% correct) compared with radiologist predictions (radiologist 1 = 67.8% correct, radiologist 2 = 68.6% correct; McNemar < .001 for both). AI-based tools have not yet reached full diagnostic potential for COVID-19 and underperform compared with radiologist prediction. Diagnosis, Classification, Application Domain, Infection, Lung . © RSNA, 2022.
ISSN:2638-6100
2638-6100
DOI:10.1148/ryai.210217