A real-world study evaluating ultrasound-guided percutaneous non-targeted liver biopsy needle failures and pathology sample-quality assessment in both end-cut and side-notch needles
To determine biopsy device failures, causative factors, complications and sample quality of the 16G end-cut Biopince and side-notch Bard needles. All ultrasound-guided non-targeted liver biopsies between 01/01/2016 and 31/12/2018 were included. Operator, device, number of failures, complications and...
Gespeichert in:
Veröffentlicht in: | British journal of radiology 2021-09, Vol.94 (1125), p.20210475-20210475 |
---|---|
Hauptverfasser: | , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To determine biopsy device failures, causative factors, complications and sample quality of the 16G end-cut Biopince
and side-notch Bard
needles.
All ultrasound-guided non-targeted liver biopsies between 01/01/2016 and 31/12/2018 were included. Operator, device, number of failures, complications and repeat biopsies were recorded. Histopathology samples were reviewed for all cases of needle failure and a group with no failures, and graded "yes/no" for the presence of steatosis, inflammation and fibrosis. The pathology slides from these cases were reviewed to assess biopsy sample quality (length and portal tract number). The failure and no-failure groups were compared in terms of device type/histology, and sample quality was compared between the needle types.
1004 patients were included. 93.8% (
= 942) required one needle pass to obtain a sample and 6.2% (
= 62) required >1 pass due to needle failure. Total of 76 needle failures, more with end-cut than side-notch needles (8.7%
2.9%) (
< 0.001). No needle failures resulted in complication. The presence of liver fibrosis was associated with fewer needle failures (
= 0.036). The major complication rate was 0.4% (4/1044). A biopsy with >10 portal tracts was obtained in 90.2% of specimens > 20 mm long, compared with 66% of 16-20 mm biopsies and 21% of 10 portal tracts was achieved in 10/26 (38.5%) of side-notch biopsies and 64/90 (71.1%) of end-cut biopsies (
= 0.004).
Ultrasound-guided liver biopsy is safe and sample quality is consistently good when a core >20 mm long is obtained. The end-cut biopsy device generated reliably good quality biopsy samples; however, the needle failure rate was significantly higher than the side-cut needle.
Ultrasound-guided liver biopsy specimen quality is consistently good when a core >20 mm long is obtained which can be achieved with a single pass using the 16G Biopince
end-cut needle, although the needle failure rate is significantly higher than the 16G Max-Core
Bard
side-notch needle. |
---|---|