A real-world study evaluating ultrasound-guided percutaneous non-targeted liver biopsy needle failures and pathology sample-quality assessment in both end-cut and side-notch needles

To determine biopsy device failures, causative factors, complications and sample quality of the 16G end-cut Biopince and side-notch Bard needles. All ultrasound-guided non-targeted liver biopsies between 01/01/2016 and 31/12/2018 were included. Operator, device, number of failures, complications and...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:British journal of radiology 2021-09, Vol.94 (1125), p.20210475-20210475
Hauptverfasser: Zhong, Jim, Allard, Rachel, Hewitson, Dominic, Weston, Michael, Hulson, Oliver, Burbidge, Simon, Lambie, Hannah, Kaye, Tom, Wyatt, Judith I, Albazaz, Raneem
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To determine biopsy device failures, causative factors, complications and sample quality of the 16G end-cut Biopince and side-notch Bard needles. All ultrasound-guided non-targeted liver biopsies between 01/01/2016 and 31/12/2018 were included. Operator, device, number of failures, complications and repeat biopsies were recorded. Histopathology samples were reviewed for all cases of needle failure and a group with no failures, and graded "yes/no" for the presence of steatosis, inflammation and fibrosis. The pathology slides from these cases were reviewed to assess biopsy sample quality (length and portal tract number). The failure and no-failure groups were compared in terms of device type/histology, and sample quality was compared between the needle types. 1004 patients were included. 93.8% ( = 942) required one needle pass to obtain a sample and 6.2% ( = 62) required >1 pass due to needle failure. Total of 76 needle failures, more with end-cut than side-notch needles (8.7% 2.9%) ( < 0.001). No needle failures resulted in complication. The presence of liver fibrosis was associated with fewer needle failures ( = 0.036). The major complication rate was 0.4% (4/1044). A biopsy with >10 portal tracts was obtained in 90.2% of specimens > 20 mm long, compared with 66% of 16-20 mm biopsies and 21% of 10 portal tracts was achieved in 10/26 (38.5%) of side-notch biopsies and 64/90 (71.1%) of end-cut biopsies ( = 0.004). Ultrasound-guided liver biopsy is safe and sample quality is consistently good when a core >20 mm long is obtained. The end-cut biopsy device generated reliably good quality biopsy samples; however, the needle failure rate was significantly higher than the side-cut needle. Ultrasound-guided liver biopsy specimen quality is consistently good when a core >20 mm long is obtained which can be achieved with a single pass using the 16G Biopince end-cut needle, although the needle failure rate is significantly higher than the 16G Max-Core Bard side-notch needle.