Comparison of Total Corneal Astigmatism between IOLMaster and Pentacam

Purpose. To compare the total corneal astigmatism (TCA) measured by IOLMaster 700 and Pentacam and to investigate the consistency of corneal keratometry (CK) measured by IOLMaster and Pentacam. Methods. Cataract patients were retrospectively enrolled in March and April, 2021. Retrospective analysis...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BioMed research international 2022-07, Vol.2022, p.1-7
Hauptverfasser: Li, Xiaochun, Cao, Xiaoguang, Bao, Yongzhen
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose. To compare the total corneal astigmatism (TCA) measured by IOLMaster 700 and Pentacam and to investigate the consistency of corneal keratometry (CK) measured by IOLMaster and Pentacam. Methods. Cataract patients were retrospectively enrolled in March and April, 2021. Retrospective analysis was performed on those patients with binocular and monocular CK measured by IOLMaster and Pentacam. Results. A total of 102 patients (204 eyes) were included, 64 of whom were female (62.75%). The flat (K1) and steep (K2) CK of anterior corneal surface (ACS) and flat (TK1) and steep (TK2) of total cornea measured with IOLMaster 700 were 44.16±1.60 D, 45.09±1.68 D, 44.12±1.62 D, and 45.14±1.69 D, respectively. Those measured with Pentacam were 44.31±1.57 D, 45.22±1.65 D, 44.15±1.67 D, and 45.19±1.82 D, respectively. The astigmatism of ACS and TCA were −0.94±0.63 D and −1.02±0.69 D (p