Accuracy Evaluation of Collapsed Cone Convolution Superposition Algorithms for the Nasopharynx Interface in the Early Stage of Nasopharyngeal Carcinoma
This study combined the use of radiation dosimeteric measurements and a custom-made anthropomorphic phantom in order to evaluate the accuracy of therapeutic dose calculations at the nasopharyngeal air-tissue interface. The doses at the nasopharyngeal air-tissue interface obtained utilizing the Pinna...
Gespeichert in:
Veröffentlicht in: | BioMed research international 2022, Vol.2022, p.5227609-13 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | This study combined the use of radiation dosimeteric measurements and a custom-made anthropomorphic phantom in order to evaluate the accuracy of therapeutic dose calculations at the nasopharyngeal air-tissue interface. The doses at the nasopharyngeal air-tissue interface obtained utilizing the Pinnacle and TomoTherapy TPS, which are based on collapsed cone convolution superposition (CCCS) algorithms, were evaluated and measured under single 10×10 cm2, 2×2 cm2, two parallel opposed 2×2 cm2 and clinical fields for early stage of nasopharyngeal carcinoma by using EBT3, GR-200F, and TLD 100. At the air-tissue interface under a 10×10 cm2 field, the TPS dose calculation values were in good agreement with the dosimeter measurement with all differences within 3.5%. When measured the single field 2×2 cm2, the differences between the average dose were measured at the distal interface for EBT3, GR-200F, and TLD-100 and the calculation values were -15.8%, -16.4%, and -4.9%, respectively. When using the clinical techniques such as IMRT, VMAT, and tomotherapy, the measurement results at the interface for all three techniques did not imply under dose. Small-field sizes will lead to dose overestimation at the nasopharyngeal air-tissue interface due to electronic disequilibrium when using CCCS algorithms. However, under clinical applications of multiangle irradiation, the dose errors caused by this effect were not significant. |
---|---|
ISSN: | 2314-6133 2314-6141 |
DOI: | 10.1155/2022/5227609 |