Possibilities of monitoring intraocular pressure in children using EASYTON transpalpebral tonometer
Purpose To compare the effectiveness of transpalpebral scleral tonometry (TPST) and corneal pneumotonometry in children, and assess the discomfort level when measuring intraocular pressure (IOP) by these methods. Methods TPST using EASYTON tonometer (Russia) and pneumotonometry using Reichert 7 Non-...
Gespeichert in:
Veröffentlicht in: | International ophthalmology 2022-05, Vol.42 (5), p.1631-1638 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To compare the effectiveness of transpalpebral scleral tonometry (TPST) and corneal pneumotonometry in children, and assess the discomfort level when measuring intraocular pressure (IOP) by these methods.
Methods
TPST using EASYTON tonometer (Russia) and pneumotonometry using Reichert 7 Non-contact AutoTonometer (USA) have been sequentially performed on 84 eyes (42 children aged 5–14, ave. 9.3 ± 2.7), including 64 myopic eyes (-0.5 to 6.75D), 18 hyperopic eyes (+ 0.75 to + 3.75D), and 2 emmetropic eyes. We assessed tolerance to the procedure on a five-point scale using a questionnaire which listed several criteria: discomfort, presence of pain, fear or anxiety during the procedure, the child's resistance to measurement.
Results
EASYTON tonometry demonstrated repeatability of IOP indicators when measuring the same eye three times sequentially and almost the same IOP level in paired eyes of isometropic children. Pneumotonometry reveals a greater individual data variability and a more pronounced asymmetry of the paired eyes’ indicators. IOP measured using the TPST was 18.3 ± 2.3 mmHg across the whole group, 18.2 ± 2.3 mmHg in myopic, and 18.5 ± 2.3 mmHg in hyperopic children. With pneumotonometry, the corresponding indicators were 17.1 ± 3.9 mmHg, 16.9 ± 3.8 mmHg, and 18.2 ± 4.0 mmHg. The average score for the TPST (4.64 ± 0.60 points) was significantly higher than that for pneumotonometry (3.85 ± 0.90 points) (
p
|
---|---|
ISSN: | 1573-2630 0165-5701 1573-2630 |
DOI: | 10.1007/s10792-021-02158-5 |