Application of digital PCR to determine the reliability of Xpert Xpress SARS-CoV-2 assay with envelope (E) gene negative and nucleocapsid (N2) gene positive results
•E gene negative & N2 gene positive (E-N+) result by Cepheid could be falsely positive.•Application of digital PCR (dPCR) to determine whether E-N+ result is reliable.•Two dPCR assays were compared in detection of SARS-CoV-2 in low viral load samples.•QX200 ddPCR confirmed 81.8% of the E-N+ samp...
Gespeichert in:
Veröffentlicht in: | Diagnostic microbiology and infectious disease 2022-08, Vol.103 (4), p.115726-115726, Article 115726 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •E gene negative & N2 gene positive (E-N+) result by Cepheid could be falsely positive.•Application of digital PCR (dPCR) to determine whether E-N+ result is reliable.•Two dPCR assays were compared in detection of SARS-CoV-2 in low viral load samples.•QX200 ddPCR confirmed 81.8% of the E-N+ samples by Cepheid were true positive.
This study used digital polymerase chain reaction (dPCR) to determine whether envelope (E) gene-negative and nucleocapsid (N2) gene-positive (E-N+) results obtained with the Cepheid Xpert Xpress SARS-CoV-2 assay are reliable. Using droplet digital PCR results as a reference, 18 of 22 E-N+ samples with a low viral load (81.8%) were identified as true positives. |
---|---|
ISSN: | 0732-8893 1879-0070 |