Clinical experience with an in-NICU magnetic resonance imaging system
Objective To evaluate the utility of the 1 Tesla (1 T) Embrace (Aspect Imaging) neonatal magnetic resonance imaging (MRI) scanner in a level III NICU. Study design Embrace brain MRI findings for 207 infants were reviewed, including 32 scans directly compared within 5 days with imaging on a 3 T Sieme...
Gespeichert in:
Veröffentlicht in: | Journal of perinatology 2022-07, Vol.42 (7), p.873-879 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objective
To evaluate the utility of the 1 Tesla (1 T) Embrace (Aspect Imaging) neonatal magnetic resonance imaging (MRI) scanner in a level III NICU.
Study design
Embrace brain MRI findings for 207 infants were reviewed, including 32 scans directly compared within 5 days with imaging on a 3 T Siemens Trio. Clinical MRI scan abnormalities were also compared to cranial ultrasound findings.
Result
Of the 207 Embrace brain MRIs, 146 (70.5%) were obtained for clinical indications and 61 (29.5%) were research cases. Abnormal findings were found in 80 scans, most commonly hemorrhage and white matter injury. Notable findings included a stroke, medullary brainstem tumor, and polymicrogyria. In the 1 T versus 3 T comparison cohort, results were discordant in only one infant with punctate foci of susceptibility noted only on the 3 T scan.
Conclusion
The Embrace MRI scans detected clinically relevant brain abnormalities and in a subset were clinically comparable to 3 T scans. |
---|---|
ISSN: | 0743-8346 1476-5543 |
DOI: | 10.1038/s41372-022-01387-5 |