Alinity hq platelet count is not impacted by severe microcytosis
Background Impedance technology has been shown to overestimate platelet (PLT) count in samples with microcytes, while the optical‐fluorescence PLT count (PLT‐F) by Sysmex has been suggested to be unaffected by microcytosis. The Abbott Alinity hq analyzer employs multi‐dimensional optical PLT countin...
Gespeichert in:
Veröffentlicht in: | Journal of clinical laboratory analysis 2022-04, Vol.36 (4), p.e24218-n/a |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
Impedance technology has been shown to overestimate platelet (PLT) count in samples with microcytes, while the optical‐fluorescence PLT count (PLT‐F) by Sysmex has been suggested to be unaffected by microcytosis. The Abbott Alinity hq analyzer employs multi‐dimensional optical PLT counting. Our goal was to assess the accuracy of this technology in microcytic samples.
Methods
Platelet measurements were performed by Alinity hq and the impedance (PLT‐I) and PLT‐F methods on a Sysmex XN‐3000 analyzer on 464 samples. PLT concentration range was 6.56–947 × 109/L and mean cell volume (MCV) 40.9–123.0 fL. Samples were categorized into normocytic (MCV > 80 fL), microcytic (MCV 65–80 fL), and severely microcytic (MCV |
---|---|
ISSN: | 0887-8013 1098-2825 |
DOI: | 10.1002/jcla.24218 |