Human Vitreous Collagen Fragments Dimension As a Function of Vitrectomy Cut Rate
To study the dimensions and distribution of human vitreous collagen type II fragments collected after vitrectomy performed at varying cut rates and to evaluate if increasing the cut rate produces smaller collagen fragments, thus reducing retinal traction and/or viscosity. Fluid was collected during...
Gespeichert in:
Veröffentlicht in: | Translational vision science & technology 2022-03, Vol.11 (3), p.29-29 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To study the dimensions and distribution of human vitreous collagen type II fragments collected after vitrectomy performed at varying cut rates and to evaluate if increasing the cut rate produces smaller collagen fragments, thus reducing retinal traction and/or viscosity.
Fluid was collected during core vitrectomies performed for macular surgery at cut rates from 1000 to 16,000 cuts per minute (CPM) and immediately refrigerated. Protein fractions were separated by molecular weight (MW; >100 kDa, 50-100 kDa, 50-30 kDa, 30-10 kDa, and |
---|---|
ISSN: | 2164-2591 2164-2591 |
DOI: | 10.1167/tvst.11.3.29 |