Three-Dimensional Analysis of the In Vivo Motion of Implantable Cardioverter Defibrillator Leads
Better understanding of the lead curvature, movement and their spatial distribution may be beneficial in developing lead testing methods, guiding implantations and improving life expectancy of implanted leads. Objective The aim of this two-phase study was to develop and test a novel biplane cine-flu...
Gespeichert in:
Veröffentlicht in: | Cardiovascular engineering and technology 2022-02, Vol.13 (1), p.129-138 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Better understanding of the lead curvature, movement and their spatial distribution may be beneficial in developing lead testing methods, guiding implantations and improving life expectancy of implanted leads.
Objective
The aim of this two-phase study was to develop and test a novel biplane cine-fluoroscopy-based method to evaluate input parameters for bending stress in leads based on their
in vivo
3D motion using precisely determined spatial distributions of lead curvatures. Potential tensile, compressive or torque forces were not subjects of this study.
Methods
A method to measure lead curvature and curvature evolution was initially tested in a phantom study. In the second phase using this model 51 patients with implanted ICD leads were included. A biplane cine-fluoroscopy recording of the intracardiac region of the lead was performed. The lead centerline and its motion were reconstructed in 3D and used to define lead curvature and curvature changes. The maximum absolute curvature
C
max
during a cardiac cycle, the maximum curvature amplitude
C
amp
and the maximum curvature
C
max@amp
at the location of
C
amp
were calculated. These parameters can be used to characterize fatigue stress in a lead under cyclical bending.
Results
The medians of
C
amp
and
C
max@amp
were 0.18 cm
−1
and 0.42 cm
−1
, respectively. The median location of
C
max
was in the atrium whereas the median location of
C
amp
occurred close to where the transit through the tricuspid valve can be assumed. Increased curvatures were found for higher slack grades.
Conclusion
Our results suggest that reconstruction of 3D ICD lead motion is feasible using biplane cine-fluoroscopy. Lead curvatures can be computed with high accuracy and the results can be implemented to improve lead design and testing. |
---|---|
ISSN: | 1869-408X 1869-4098 |
DOI: | 10.1007/s13239-021-00557-4 |