Evaluation of Accuracy and Safety of the Next-Generation Up to 180-Day Long-Term Implantable Eversense Continuous Glucose Monitoring System: The PROMISE Study
Use of continuous glucose monitoring (CGM) systems is being rapidly adopted as standard of care for insulin-requiring patients with diabetes. The PROMISE study (NCT03808376) evaluated the accuracy and safety of the next-generation implantable Eversense CGM system for up to 180 days. This was a prosp...
Gespeichert in:
Veröffentlicht in: | Diabetes technology & therapeutics 2022-02, Vol.24 (2), p.84-92 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Use of continuous glucose monitoring (CGM) systems is being rapidly adopted as standard of care for insulin-requiring patients with diabetes. The PROMISE study (NCT03808376) evaluated the accuracy and safety of the next-generation implantable Eversense CGM system for up to 180 days.
This was a prospective multicenter study involving 181 subjects with diabetes at 8 USA sites. All subjects were inserted with a primary sensor. Ninety-six subjects had a second sensor, either an identical sensor or a modified sensor (sacrificial boronic acid [SBA]), inserted in their other arm (53 and 43 subjects, respectively). Accuracy was evaluated by comparing CGM to YSI 2300 glucose analyzer (Yellow Springs Instrument [YSI]) values during 10 clinic visits (day 1-180). Confirmed event detection rates, calibration stability, sensor survival, and serious adverse events (SAEs) were evaluated.
For primary sensors, the percent CGM readings within 20%/20% of YSI values was 92.9%; overall mean absolute relative difference (MARD) was 9.1%. The confirmed alert detection rate at 70 mg/dL was 93% and at 180 mg/dL was 99%. The median percentage of time for one calibration per day was 56%. Sixty-five percent of the primary sensors survived to 180 days. For the SBA sensors, the percent CGM readings within 20%/20% of YSI values was 93.9%; overall MARD was 8.5%. The confirmed alert detection rate at 70 mg/dL was 94% and at 180 mg/dL was 99%. The median percentage of time for one calibration per day was 63%. Ninety percent of the SBA sensors survived to 180 days. No device- or insertion/removal procedure-related SAEs were reported.
These data show the next-generation Eversense CGM system had sustained accuracy and safety up to 180 days, with an improved calibration scheme and survival, using the primary or SBA sensors. |
---|---|