Multicentre experience with an iliac fenestrated device
Abstract OBJECTIVES The aim of this study was to present a multicentre experience of technical results and mid-term follow-up using a custom-made iliac fenestrated device (Terumo Aortic, Inchinnan, Renfrewshire, UK) for the treatment of iliac aneurysms and endoleaks. METHODS A multicentre retrospe...
Gespeichert in:
Veröffentlicht in: | Interactive cardiovascular and thoracic surgery 2021-08, Vol.33 (3), p.448-454 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext bestellen |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
OBJECTIVES
The aim of this study was to present a multicentre experience of technical results and mid-term follow-up using a custom-made iliac fenestrated device (Terumo Aortic, Inchinnan, Renfrewshire, UK) for the treatment of iliac aneurysms and endoleaks.
METHODS
A multicentre retrospective evaluation of 22 patients (3–12 per institution) with either an iliac artery aneurysm or endoleak treated with an iliac fenestrated device was performed. Data were gathered from 3 departments of vascular and endovascular surgery at 3 European institutions.
RESULTS
Ten of the included patients (45.5%) were treated for an endoleak and 12 had aorto-iliac aneurysms (54.5%). Two patients underwent bilateral fenestrated device implantation for a total of 24 devices included in this analysis. Primary technical success was 91.7% (22 of 24 implanted devices). One of the 24 internal iliac arteries could not be cannulated and was covered (primary assisted technical success rate 95.8%) and 1 patient required a relining of the stent graft due to a mid-grade stenosis opposite the internal iliac artery fenestration. Survival at the last available follow-up (mean 15.2 ± 12.0 months, range 0.5–36.6 months) was 90.9%.
CONCLUSIONS
The present investigation adds to a growing body of literature on custom-made endografts and their usefulness in achieving endovascular repair without compromising blood flow via important arterial branch vessels, such as the internal iliac artery. It presents encouraging technical and mid-term follow-up data from consecutive patients treated for iliac aneurysms or endoleaks using this custom-made device. The technique may help avoid adverse sequelae associated to a coil-and-cover approach when iliac branch devices are not feasible. |
---|---|
ISSN: | 1569-9285 1569-9293 1569-9285 |
DOI: | 10.1093/icvts/ivab115 |