Total Retention Liquid Chromatography–Mass Spectrometry to Achieve Maximum Protein Sequence Coverage
Peptide identification by liquid chromatography–mass spectrometry (LC–MS) requires retention and elution of peptides from the LC column. Although medium and hydrophobic peptides are readily retained by the C18 columns that are commonly used in proteomics, short and hydrophilic peptides are not retai...
Gespeichert in:
Veröffentlicht in: | Analytical chemistry (Washington) 2021-03, Vol.93 (12), p.5054-5060 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Peptide identification by liquid chromatography–mass spectrometry (LC–MS) requires retention and elution of peptides from the LC column. Although medium and hydrophobic peptides are readily retained by the C18 columns that are commonly used in proteomics, short and hydrophilic peptides are not retained nor measured by MS due to their elution in the void volume after sample injection. These nonretained peptides can possess important post-translational modifications, such as glycosylation or phosphorylation. We describe a total retention LC–MS method that employs a reverse phase C18 column and porous graphitic carbon (PGC) column to retain both hydrophobic and hydrophilic peptides for LC–MS analysis. Our setup uses a single valve with a trapping column and two LC pumps run at low microliter/minute flow rates to deliver separate gradients to parallel capillary C18 and PGC columns. Our capillary LC system balances the need for high sensitivity with ease of implementation as compared to other 2D LC systems that use nanocolumns with multiple trapping columns and multiport valves. We demonstrate the utility of the method identifying hydrophilic peptides that went undetected when only a C18 nanocolumn was used. These missed hydrophilic peptides include tripeptides and N-glycosylated species. |
---|---|
ISSN: | 0003-2700 1520-6882 |
DOI: | 10.1021/acs.analchem.0c04292 |