A Novel Proton Pencil Beam Scanning FLASH RT Delivery Method Enables Optimal OAR Sparing and Ultra-High Dose Rate Delivery: A Comprehensive Dosimetry Study for Lung Tumors

While transmission proton beams have been demonstrated to achieve ultra-high dose rate FLASH therapy delivery, they are unable to spare normal tissues distal to the target. This study aims to compare FLASH treatment planning using single energy Bragg peak proton beams versus transmission proton beam...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Cancers 2021-11, Vol.13 (22), p.5790
Hauptverfasser: Wei, Shouyi, Lin, Haibo, Choi, J Isabelle, Simone, 2nd, Charles B, Kang, Minglei
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:While transmission proton beams have been demonstrated to achieve ultra-high dose rate FLASH therapy delivery, they are unable to spare normal tissues distal to the target. This study aims to compare FLASH treatment planning using single energy Bragg peak proton beams versus transmission proton beams in lung tumors and to evaluate Bragg peak plan optimization, characterize plan quality, and quantify organ-at-risk (OAR) sparing. Both Bragg peak and transmission plans were optimized using an in-house platform for 10 consecutive lung patients previously treated with proton stereotactic body radiation therapy (SBRT). To bring the dose rate up to the FLASH-RT threshold, Bragg peak plans with a minimum MU/spot of 1200 and transmission plans with a minimum MU/spot of 400 were developed. Two common prescriptions, 34 Gy in 1 fraction and 54 Gy in 3 fractions, were studied with the same beam arrangement for both Bragg peak and transmission plans ( = 40 plans). RTOG 0915 dosimetry metrics and dose rate metrics based on different dose rate calculations, including average dose rate (ADR), dose-averaged dose rate (DADR), and dose threshold dose rate (DTDR), were investigated. We then evaluated the effect of beam angular optimization on the Bragg peak plans to explore the potential for superior OAR sparing. Bragg peak plans significantly reduced doses to several OAR dose parameters, including lung V and V by 32.0% ( < 0.01) and 30.4% ( < 0.01) for 34Gy/fx plans, respectively; and by 40.8% ( < 0.01) and 41.2% ( < 0.01) for 18Gy/fx plans, respectively, compared with transmission plans. Bragg peak plans have ~3% less in DADR and ~10% differences in mean OARs in DTDR and DADR relative to transmission plans due to the larger portion of lower dose regions of Bragg peak plans. With angular optimization, optimized Bragg peak plans can further reduce the lung V by 20.7% ( < 0.01) and V by 19.7% ( < 0.01) compared with Bragg peak plans without angular optimization while achieving a similar 3D dose rate distribution. The single-energy Bragg peak plans achieve superior dosimetry performances in OARs to transmission plans with comparable dose rate performances for lung cancer FLASH therapy. Beam angle optimization can further improve the OAR dosimetry parameters with similar 3D FLASH dose rate coverage.
ISSN:2072-6694
2072-6694
DOI:10.3390/cancers13225790