Adverse events related to total ankle replacement devices: an analysis of reports to the United States Food and Drug Administration
Background The published outcomes of total ankle replacement (TAR) implants came from limited institutions creating observational bias. For broader perspective, we queried the Food and Drug Administration’s (FDA) Manufacturer and User Facility Device Experience (MAUDE) voluntary database to explore...
Gespeichert in:
Veröffentlicht in: | International orthopaedics 2021-09, Vol.45 (9), p.2307-2312 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The published outcomes of total ankle replacement (TAR) implants came from limited institutions creating observational bias. For broader perspective, we queried the Food and Drug Administration’s (FDA) Manufacturer and User Facility Device Experience (MAUDE) voluntary database to explore complications reported outside published literature.
Methods
The database was reviewed retrospectively between November 2011 and April 2019 using two product codes assigned to six TAR devices.
Results
Among 648 relevant reports available in the database, common complications were aseptic loosening (19.3%), infection (18.2%), and alignment/mechanical issues (16.5%). Others included instrument/instrumentation complications, impingement, polyethylene problems, fractures, avascular necrosis of talus (AVN), and packaging issues.
Conclusion
MAUDE database revealed various patterns of device-related malfunctions that have been under-reported in published data. Despite inconsistency in the available reports, it provided opportunities for improvements in quality control, device design, and ultimately patient safety. Database would be further strengthened by more robust reporting mechanism or mandatory reporting of device-related complications. |
---|---|
ISSN: | 0341-2695 1432-5195 |
DOI: | 10.1007/s00264-021-04972-z |