Risk Management of Clinical Reference Dosimetry of a Large Hospital Network Using Statistical Process Control
Managing TG-51 reference dosimetry in a large hospital network can be a challenging task. The objectives of this study are to investigate the effectiveness of using Statistical Process Control (SPC) to manage TG-51 workflow in such a network. All the sites in the network performed the annual referen...
Gespeichert in:
Veröffentlicht in: | International journal of medical physics, clinical engineering and radiation oncology clinical engineering and radiation oncology, 2021-08, Vol.10 (3), p.119-131 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Managing TG-51 reference dosimetry in a large hospital network can be a challenging task. The objectives of this study are to investigate the effectiveness of using Statistical Process Control (SPC) to manage TG-51 workflow in such a network. All the sites in the network performed the annual reference dosimetry in water according to TG-51. These data were used to cross-calibrate the same ion chambers in plastic phantoms for monthly QA output measurements. An energy-specific dimensionless beam quality cross-calibration factor,
k
q
n
S
W
, was derived to monitor the process across multiple sites. The SPC analysis was then performed to obtain the mean,
〈
k
q
n
S
W
〉
, standard deviation,
σ
k
, the Upper Control Limit (UCL) and Lower Control Limit (LCL) in each beam. This process was first applied to 15 years of historical data at the main campus to assess the effectiveness of the process. A two-year prospective study including all 30 linear accelerators spread over the main campus and seven satellites in the network followed. The ranges of the control limits (±3
σ
) were found to be in the range of 1.7% – 2.6% and 3.3% – 4.2% for the main campus and the satellite sites respectively. The wider range in the satellite sites was attributed to variations in the workflow. Standardization of workflow was also found to be effective in narrowing the control limits. The SPC is effective in identifying variations in the workflow and was shown to be an effective tool in managing large network reference dosimetry. |
---|---|
ISSN: | 2168-5436 2168-5444 |
DOI: | 10.4236/ijmpcero.2021.103011 |