Fast spin-echo MR of contact points on implanted intracerebral stainless steel multicontact electrodes

A three-dimensional fast spin-echo MR technique is proposed for locating contact points on implanted intracerebral multicontact electrode bundles, Coronal or sagittal reformatting shows the entire trajectory of the electrode bundles. The contract points are clearly visible owing to the absence of co...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of neuroradiology : AJNR 1996-11, Vol.17 (10), p.1815-1819
Hauptverfasser: Meiners, LC, Bakker, CJ, van Rijen, PC, van Veelen, CW, van Huffelen, AC, van Dieren, A, Jansen, GH, Mali, WP
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A three-dimensional fast spin-echo MR technique is proposed for locating contact points on implanted intracerebral multicontact electrode bundles, Coronal or sagittal reformatting shows the entire trajectory of the electrode bundles. The contract points are clearly visible owing to the absence of coating material associated with a slightly larger susceptibility artifact. Potentially, this technique may preclude postimplantation thin-section CT, with its associated high radiation dose.
ISSN:0195-6108
1936-959X