Smartphone Inclinometry Is a Valid and Reliable Tool for Measuring Frontal Plane Tibial Alignment in Healthy and Osteoarthritic Knees
Abstract Objective Frontal plane knee alignment plays an integral role in tibiofemoral knee osteoarthritis development and progression. Accessible methods for obtaining direct or indirect measures of knee alignment may help inform clinical decision making when specialized equipment is unavailable. T...
Gespeichert in:
Veröffentlicht in: | PTJ: Physical Therapy & Rehabilitation Journal 2021-07, Vol.101 (7), p.1 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Abstract
Objective
Frontal plane knee alignment plays an integral role in tibiofemoral knee osteoarthritis development and progression. Accessible methods for obtaining direct or indirect measures of knee alignment may help inform clinical decision making when specialized equipment is unavailable. The present study evaluated the concurrent validity, as well as intersession (within-rater) and interrater (within-session) reliability of smartphone inclinometry for measuring static frontal plane tibial alignment—a known proxy of frontal plane knee alignment.
Methods
Twenty healthy individuals and 38 patients with knee osteoarthritis were measured for frontal plane tibial alignment by a pair of raters using smartphone inclinometry, manual inclinometry, and 3-dimensional motion capture simultaneously. Healthy participants were measured on 2 separate days. Bland-Altman analysis, supplemented with intraclass correlation coefficient (ICC)(2,k), was used to assess concurrent validity. ICC(2,k), SEM, and minimum detectable change with 95% confidence limits (MDC95) were used to assess measurement reliability.
Results
Compared against motion capture, smartphone inclinometry measured frontal plane tibial alignment with a mean difference of 0.7 and 1.1 degrees (biased toward varus) for healthy participants and participants with knee osteoarthritis, respectively (ICC[2,k] ≥ 0.87). Smartphone inclinometry measurements demonstrated adequate intersession (within-rater) relative (ICC[2,k] = 0.91) and absolute (SEM = 0.7 degrees; MDC95 = 1.8 degrees) reliability, which outperformed manual inclinometry (ICC[2,k] = 0.85; SEM = 1.0 degrees; MDC95 = 2.6 degrees). Interrater (within-session) reliability of smartphone inclinometry was acceptable in both cohorts (ICC[2,k] = 0.93; SEM = 0.4 degrees to 1.2 degrees; MDC95 = 1.2 degrees to 3.2 degrees).
Conclusion
Smartphone inclinometry is sufficiently valid and reliable for measuring frontal plane tibial alignment in healthy individuals and patients with medial tibiofemoral knee osteoarthritis.
Impact
Smartphones are readily accessible by clinicians and researchers. Our assessment of measurement validity and reliability supports the use of smartphone inclinometry as a clinically available tool to measure frontal plane tibial alignment without medical imaging or specialized equipment. |
---|---|
ISSN: | 0031-9023 1538-6724 1538-6724 |
DOI: | 10.1093/ptj/pzab091 |