Renal Protection Phenomenon Observed in a Porcine Model After Electromagnetic Lithotripsy Using a Treatment Pause

Pretreating the kidney with 100 low-energy shock waves (SWs) with a time pause before delivering a clinical dose of SWs (Dornier HM-3, 2000 SWs, 24 kV, and 120 SWs/min) has been shown to significantly reduce the size of the hemorrhagic lesion produced in that treated kidney, compared to a protocol w...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of endourology 2021-05, Vol.35 (5), p.682-686
Hauptverfasser: Connors, Bret A, Gardner, Tony, Liu, Ziyue, Lingeman, James E, Williams, Jr, James C
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Pretreating the kidney with 100 low-energy shock waves (SWs) with a time pause before delivering a clinical dose of SWs (Dornier HM-3, 2000 SWs, 24 kV, and 120 SWs/min) has been shown to significantly reduce the size of the hemorrhagic lesion produced in that treated kidney, compared to a protocol without pretreatment. It has been assumed that a similar reduction in injury will occur with lithotripters other than the HM-3, but experiments to confirm this assumption are lacking. In this study, we sought to verify that the lesion protection phenomenon also occurs in a lithotripter using an electromagnetic shock source and dry-head coupling. Eleven female pigs were placed in a Dornier Compact S lithotripter where the left kidney of each animal was targeted for lithotripsy treatment. Some kidneys received 2500 SWs at power level (PL) = 5 (120 SWs/min) while some kidneys were pretreated with 100 SWs at PL = 1, with a 3-minute time pause, followed immediately by 2500 SWs at PL = 5 (120 SWs/min). Lesion size analysis was performed to assess the volume of hemorrhagic tissue injury created by each treatment regimen (% functional renal volume). Lesion size fell by 85% (  = 0.01) in the 100 SW pretreatment group compared to the injury produced by a regimen without pretreatment. The results suggest that the treatment pause protection phenomenon occurs with a Dornier Compact S, a machine distinctly different from the Dornier HM-3. This result also suggests that this phenomenon may be observed generally in SW lithotripters.
ISSN:0892-7790
1557-900X
DOI:10.1089/end.2020.0681