Effect of Orthodontic Brackets on the Accuracy of Apex Locators: A Pilot Study

The purpose of this study was to evaluate the effect of metal orthodontic brackets on the accuracy of electronic apex locator (EAL). The actual canal lengths (ACL) of 40 mandibular incisor teeth were determined. Then, the teeth were randomly divided into two groups (n=20). Orthodontic metal brackets...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BioMed research international 2021, Vol.2021, p.6615560-5
Hauptverfasser: Şen, Özgür Genç, Ulusoy, Özgür İlke, Paltun, Yelda Nayır, Ulusoy, Çağrı
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The purpose of this study was to evaluate the effect of metal orthodontic brackets on the accuracy of electronic apex locator (EAL). The actual canal lengths (ACL) of 40 mandibular incisor teeth were determined. Then, the teeth were randomly divided into two groups (n=20). Orthodontic metal brackets were applied in the first group, and no brackets, in the second group. The working length of each tooth was measured with an EAL under 3 test conditions according to the distance between the lip clip and sample tooth. Data were analyzed using one-way repeated measures analysis of variance and Tukey’s post hoc tests (p=0.05). In the bracketed samples, when the lip clip was located at 1 cm and 2 cm from the samples. The mean differences between the EAL measurements and ACLs were statistically higher than those when the samples were located 3 cm from the lip clip (p
ISSN:2314-6133
2314-6141