Performance of the Abbott ID NOW rapid SARS-CoV-2 amplification assay in relation to nasopharyngeal viral RNA loads
•Calibration of the Xpert with viral stock dilutions was linear and reproducible.•Overall Xpert/ID-NOW positive agreement = 70% and negative agreement = 99.6%.•Sensitivity of ID-NOW for Xpert N2 CT ≤ 37 (viral loads ≥ 945 copies/mL) was 100%.•Sensitivity of ID-NOW for Xpert N2 for N2 CT > 37 was...
Gespeichert in:
Veröffentlicht in: | Journal of clinical virology 2021-07, Vol.140, p.104843-104843, Article 104843 |
---|---|
Hauptverfasser: | , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Calibration of the Xpert with viral stock dilutions was linear and reproducible.•Overall Xpert/ID-NOW positive agreement = 70% and negative agreement = 99.6%.•Sensitivity of ID-NOW for Xpert N2 CT ≤ 37 (viral loads ≥ 945 copies/mL) was 100%.•Sensitivity of ID-NOW for Xpert N2 for N2 CT > 37 was only 37.7%. |
---|---|
ISSN: | 1386-6532 1873-5967 1873-5967 |
DOI: | 10.1016/j.jcv.2021.104843 |