Extracorporeal multiorgan support including CO2-removal with the ADVanced Organ Support (ADVOS) system for COVID-19: A case report
A substantial part of COVID-19-patients suffers from multi-organ failure (MOF). We report on an 80-year old patient with pulmonary, renal, circulatory, and hepatic failure. We decided against the use of extracorporeal membrane oxygenation (ECMO) due to old age and a SOFA-score of 13. However, the pa...
Gespeichert in:
Veröffentlicht in: | International journal of artificial organs 2021-04, Vol.44 (4), p.288-294 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A substantial part of COVID-19-patients suffers from multi-organ failure (MOF). We report on an 80-year old patient with pulmonary, renal, circulatory, and hepatic failure. We decided against the use of extracorporeal membrane oxygenation (ECMO) due to old age and a SOFA-score of 13. However, the patient was continuously treated with the extracorporeal multi-organ- “ADVanced Organ Support” (ADVOS) device (ADVITOS GmbH, Munich, Germany). During eight 24h-treatment-sessions blood flow (100–300 mL/min), dialysate flow (160–320 mL/min) and dialysate pH (7.6–9.0) were adapted to optimize arterial PaCO2 and pH. Effective CO2 removal and correction of acidosis could be demonstrated by mean arterial- versus post-dialyzer values of pCO2 (68.7 ± 13.8 vs. 26.9 ± 11.6 mmHg; p |
---|---|
ISSN: | 0391-3988 1724-6040 |
DOI: | 10.1177/0391398820961781 |