Carotid to femoral pulse wave velocity (PWV) is associated with an increase in cardiovascular morbidity and all‐cause mortality. Noninvasive approach has made this method applicable for the examination of larger populations. This study aimed to obtain reference values of PWV measured with the Vicord...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:The journal of clinical hypertension (Greenwich, Conn.) Conn.), 2013-03, Vol.15 (3), p.176-179
Hauptverfasser: Müller, Jan, Oberhoffer, Renate, Barta, Christiane, Hulpke‐Wette, Martin, Hager, Alfred
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Carotid to femoral pulse wave velocity (PWV) is associated with an increase in cardiovascular morbidity and all‐cause mortality. Noninvasive approach has made this method applicable for the examination of larger populations. This study aimed to obtain reference values of PWV measured with the Vicorder device. PWV was obtained using the oscillometric Vicorder in 318 healthy, normotensive patients (165 women, 28.7±17.6 years, range 6–83 years). A plethysmographic sensor was placed over the right carotid region to pick up the carotid pulse wave and a blood pressure cuff was placed around the upper thigh to trace the femoral pulse wave. Path length was defined as the distance from the suprasternal notch to the top of the thigh cuff. Mean PWV was 6.1±1.4 m/s and significantly increased with age (r=.842; P