Late dislocation of the capsular bag-intraocular lens-modified capsular tension ring complex after knotless transscleral suturing using 9-0 polypropylene
We report a case of late breakage of a 9-0 polypropylene transscleral suture used for fixation of a dislocated capsular bag-intraocular lens-modified capsular tension ring complex in a 52-year-old woman with Marfan syndrome. Breakage occurred despite use of a cow-hitch technique for external and int...
Gespeichert in:
Veröffentlicht in: | Digital journal of ophthalmology 2020-06, Vol.26 (2), p.7-16 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We report a case of late breakage of a 9-0 polypropylene transscleral suture used for fixation of a dislocated capsular bag-intraocular lens-modified capsular tension ring complex in a 52-year-old woman with Marfan syndrome. Breakage occurred despite use of a cow-hitch technique for external and internal fixation. We believe breakage was caused by the suture chafing on the sharp edges of the modified capsular tension ring eyelet. Cross-sectional analysis of Malyugin-modified capsular tension rings from two different manufacturers revealed a difference with respect to radius of curvature. Suturing intraocular implants with relatively sharp edges may cause suture breakage; further studies are needed to identify the critical parameters for the surface quality of sutured intraocular implants. |
---|---|
ISSN: | 1542-8958 1542-8958 |
DOI: | 10.5693/djo.02.2020.04.001 |