Eye Tracking Supported Human Factors Testing Improving Patient Training

The handling of left ventricular assist devices (LVADs) can be challenging for patients and requires appropriate training. The devices’ usability impacts patients’ safety and quality of life. In this study, an eye tracking supported human factors testing was performed to reveal problems during use a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of medical systems 2021-05, Vol.45 (5), p.55-55, Article 55
Hauptverfasser: Weiss, Kerrin Elisabeth, Hoermandinger, Christoph, Mueller, Marcus, Schmid Daners, Marianne, Potapov, Evgenij V., Falk, Volkmar, Meboldt, Mirko, Lohmeyer, Quentin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The handling of left ventricular assist devices (LVADs) can be challenging for patients and requires appropriate training. The devices’ usability impacts patients’ safety and quality of life. In this study, an eye tracking supported human factors testing was performed to reveal problems during use and test the trainings’ effectiveness. In total 32 HeartWare HVAD patients (including 6 pre-VAD patients) and 3 technical experts as control group performed a battery change (BC) and a controller change (CC) as an everyday and emergency scenario on a training device. By tracking the patients’ gaze point, task duration and pump-off time were evaluated. Patients with LVAD support ≥1 year showed significantly shorter BC task duration than patients with LVAD support