Real‐time device tracking under MRI using an acousto‐optic active marker

Purpose This work aims to demonstrate the use of an “active” acousto‐optic marker with enhanced visibility and reduced radiofrequency (RF) ‐induced heating for interventional MRI. Methods The acousto‐optic marker was fabricated using bulk piezoelectric crystal and π‐phase shifted fiber Bragg grating...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Magnetic resonance in medicine 2021-05, Vol.85 (5), p.2904-2914
Hauptverfasser: Yaras, Yusuf S., Yildirim, Dursun Korel, Herzka, Daniel A., Rogers, Toby, Campbell‐Washburn, Adrienne E., Lederman, Robert J., Degertekin, F. Levent, Kocaturk, Ozgur
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Purpose This work aims to demonstrate the use of an “active” acousto‐optic marker with enhanced visibility and reduced radiofrequency (RF) ‐induced heating for interventional MRI. Methods The acousto‐optic marker was fabricated using bulk piezoelectric crystal and π‐phase shifted fiber Bragg grating (FBGs) and coupled to a distal receiver coil on an 8F catheter. The received MR signal is transmitted over an optical fiber to mitigate RF‐induced heating. A photodetector converts the optical signal into electrical signal, which is used as the input signal to the MRI receiver plug. Acousto‐optic markers were characterized in phantom studies. RF‐induced heating risk was evaluated according to ASTM 2182 standard. In vivo real‐time tracking capability was tested in an animal model under a 0.55T scanner. Results Signal‐to‐noise ratio (SNR) levels suitable for real‐time tracking were obtained by using high sensitivity FBG and piezoelectric transducer with resonance matched to Larmor frequency. Single and multiple marker coils integrated to 8F catheters were readout for position and orientation tracking by a single acousto‐optic sensor. RF‐induced heating was significantly reduced compared to a coax cable connected reference marker. Real‐time distal tip tracking of an active device was demonstrated in an animal model with a standard real‐time cardiac MR sequence. Conclusion Acousto‐optic markers provide sufficient SNR with a simple structure for real‐time device tracking. RF‐induced heating is significantly reduced compared to conventional active markers. Also, multiple RF receiver coils connected on an acousto‐optic modulator can be used on a single catheter for determining catheter orientation and shape.
ISSN:0740-3194
1522-2594
DOI:10.1002/mrm.28625