Atrio‐ventricular synchronous pacing with a single chamber leadless pacemaker: Programming and trouble shooting for common clinical scenarios
Micra leadless pacemaker has progressed from a single chamber pacemaker that can deliver VVIR pacing to a pacing device that can provide atrio‐ventricular (AV) synchrony via a unique pacing algorithm that relies on identifying mechanical atrial contraction. This novel algorithm has its own limitatio...
Gespeichert in:
Veröffentlicht in: | Journal of cardiovascular electrophysiology 2021-02, Vol.32 (2), p.533-539 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Micra leadless pacemaker has progressed from a single chamber pacemaker that can deliver VVIR pacing to a pacing device that can provide atrio‐ventricular (AV) synchrony via a unique pacing algorithm that relies on identifying mechanical atrial contraction. This novel algorithm has its own limitations and intricacies. In this paper, we review this algorithm, suggest steps for troubleshooting and programming these devices and provide clinical examples of Micra AV cases that required changes in programming for adequate tracking of atrial activity. |
---|---|
ISSN: | 1045-3873 1540-8167 |
DOI: | 10.1111/jce.14807 |