A case of cardiac strangulation following epicardial pacemaker implantation
An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further st...
Gespeichert in:
Veröffentlicht in: | General thoracic and cardiovascular surgery 2020-12, Vol.68 (12), p.1499-1502 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 1502 |
---|---|
container_issue | 12 |
container_start_page | 1499 |
container_title | General thoracic and cardiovascular surgery |
container_volume | 68 |
creator | Miyagi, Chihiro Ochiai, Yoshie Ando, Yusuke Hisahara, Manabu Baba, Hironori Takigawa, Tomoya Tokunaga, Shigehiko |
description | An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further studies including computed tomography and cardiac catheterization revealed that the pacemaker lead had completely encircled the cardiac silhouette and was in a state of “cardiac strangulation”. We removed the previous pacing leads and generator and implanted a new epicardial dual-chamber pacing system in the right atrium and right ventricle. Additionally, an expanded polytetrafluoroethylene sheet was placed between the new leads and the heart to prevent recurrence of cardiac strangulation. |
doi_str_mv | 10.1007/s11748-020-01337-y |
format | Article |
fullrecord | <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_7680322</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2918740660</sourcerecordid><originalsourceid>FETCH-LOGICAL-c564t-4631ed2ecf0083e89b763ced3c7b8e07be164f765c4291627e4967f46df4c52c3</originalsourceid><addsrcrecordid>eNp9kU9v1DAQxS1URMvCF-gBReqFS2D8J7ZzqVStSkFU4gJny-tMFm-TONhJ0X573KZsC4eextL85vk9PUJOKXygAOpjolQJXQKDEijnqty_ICdUS15KRfnR4Q3VMXmd0g6gkppWr8gxZ0yB4OqEfL0onE1YhDbP2HjrijRFO2znzk4-DEUbui789sO2wNEvSFeM1mFvbzAWvh87O0z37BvysrVdwrcPc0V-fLr8vv5cXn-7-rK-uC5dJcVUCskpNgxdC6A56nqjJHfYcKc2GkFtkErRKlk5wWoqmUJRS9UK2bTCVczxFTlfdMd502PjcMiOOzNG39u4N8F68-9m8D_NNtwaJTXk6Fng_YNADL9mTJPpfXLY5SQY5mQY1xqA6brO6Nl_6C7MccjxTDanlQApIVNsoVwMKUVsD2YomLuuzNKVyV2Z-67MPh-9exrjcPK3nAzwBUh5NWwxPv79jOwf5CmgzQ</addsrcrecordid><sourcetype>Open Access Repository</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2918740660</pqid></control><display><type>article</type><title>A case of cardiac strangulation following epicardial pacemaker implantation</title><source>MEDLINE</source><source>SpringerLink Journals - AutoHoldings</source><source>ProQuest Central</source><creator>Miyagi, Chihiro ; Ochiai, Yoshie ; Ando, Yusuke ; Hisahara, Manabu ; Baba, Hironori ; Takigawa, Tomoya ; Tokunaga, Shigehiko</creator><creatorcontrib>Miyagi, Chihiro ; Ochiai, Yoshie ; Ando, Yusuke ; Hisahara, Manabu ; Baba, Hironori ; Takigawa, Tomoya ; Tokunaga, Shigehiko</creatorcontrib><description>An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further studies including computed tomography and cardiac catheterization revealed that the pacemaker lead had completely encircled the cardiac silhouette and was in a state of “cardiac strangulation”. We removed the previous pacing leads and generator and implanted a new epicardial dual-chamber pacing system in the right atrium and right ventricle. Additionally, an expanded polytetrafluoroethylene sheet was placed between the new leads and the heart to prevent recurrence of cardiac strangulation.</description><identifier>ISSN: 1863-6705</identifier><identifier>EISSN: 1863-6713</identifier><identifier>DOI: 10.1007/s11748-020-01337-y</identifier><identifier>PMID: 32270437</identifier><language>eng</language><publisher>Singapore: Springer Singapore</publisher><subject>Age ; Cardiac arrhythmia ; Cardiac Pacing, Artificial ; Cardiac Surgery ; Cardiology ; Case Report ; Catheters ; Child ; Coronary vessels ; Heart ; Heart Atria ; Heart Block - therapy ; Heart Ventricles ; Humans ; Male ; Medical imaging ; Medicine ; Medicine & Public Health ; Pacemaker, Artificial - adverse effects ; Pacemakers ; Pericardium ; Pulmonary arteries ; Strangulation ; Surgical Oncology ; Thoracic Surgery ; Tomography ; Veins & arteries</subject><ispartof>General thoracic and cardiovascular surgery, 2020-12, Vol.68 (12), p.1499-1502</ispartof><rights>The Author(s) 2020</rights><rights>The Author(s) 2020. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c564t-4631ed2ecf0083e89b763ced3c7b8e07be164f765c4291627e4967f46df4c52c3</citedby><cites>FETCH-LOGICAL-c564t-4631ed2ecf0083e89b763ced3c7b8e07be164f765c4291627e4967f46df4c52c3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://link.springer.com/content/pdf/10.1007/s11748-020-01337-y$$EPDF$$P50$$Gspringer$$Hfree_for_read</linktopdf><linktohtml>$$Uhttps://www.proquest.com/docview/2918740660?pq-origsite=primo$$EHTML$$P50$$Gproquest$$H</linktohtml><link.rule.ids>230,314,776,780,881,21367,27901,27902,33721,33722,41464,42533,43781,51294</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/32270437$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Miyagi, Chihiro</creatorcontrib><creatorcontrib>Ochiai, Yoshie</creatorcontrib><creatorcontrib>Ando, Yusuke</creatorcontrib><creatorcontrib>Hisahara, Manabu</creatorcontrib><creatorcontrib>Baba, Hironori</creatorcontrib><creatorcontrib>Takigawa, Tomoya</creatorcontrib><creatorcontrib>Tokunaga, Shigehiko</creatorcontrib><title>A case of cardiac strangulation following epicardial pacemaker implantation</title><title>General thoracic and cardiovascular surgery</title><addtitle>Gen Thorac Cardiovasc Surg</addtitle><addtitle>Gen Thorac Cardiovasc Surg</addtitle><description>An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further studies including computed tomography and cardiac catheterization revealed that the pacemaker lead had completely encircled the cardiac silhouette and was in a state of “cardiac strangulation”. We removed the previous pacing leads and generator and implanted a new epicardial dual-chamber pacing system in the right atrium and right ventricle. Additionally, an expanded polytetrafluoroethylene sheet was placed between the new leads and the heart to prevent recurrence of cardiac strangulation.</description><subject>Age</subject><subject>Cardiac arrhythmia</subject><subject>Cardiac Pacing, Artificial</subject><subject>Cardiac Surgery</subject><subject>Cardiology</subject><subject>Case Report</subject><subject>Catheters</subject><subject>Child</subject><subject>Coronary vessels</subject><subject>Heart</subject><subject>Heart Atria</subject><subject>Heart Block - therapy</subject><subject>Heart Ventricles</subject><subject>Humans</subject><subject>Male</subject><subject>Medical imaging</subject><subject>Medicine</subject><subject>Medicine & Public Health</subject><subject>Pacemaker, Artificial - adverse effects</subject><subject>Pacemakers</subject><subject>Pericardium</subject><subject>Pulmonary arteries</subject><subject>Strangulation</subject><subject>Surgical Oncology</subject><subject>Thoracic Surgery</subject><subject>Tomography</subject><subject>Veins & arteries</subject><issn>1863-6705</issn><issn>1863-6713</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2020</creationdate><recordtype>article</recordtype><sourceid>C6C</sourceid><sourceid>EIF</sourceid><sourceid>BENPR</sourceid><recordid>eNp9kU9v1DAQxS1URMvCF-gBReqFS2D8J7ZzqVStSkFU4gJny-tMFm-TONhJ0X573KZsC4eextL85vk9PUJOKXygAOpjolQJXQKDEijnqty_ICdUS15KRfnR4Q3VMXmd0g6gkppWr8gxZ0yB4OqEfL0onE1YhDbP2HjrijRFO2znzk4-DEUbui789sO2wNEvSFeM1mFvbzAWvh87O0z37BvysrVdwrcPc0V-fLr8vv5cXn-7-rK-uC5dJcVUCskpNgxdC6A56nqjJHfYcKc2GkFtkErRKlk5wWoqmUJRS9UK2bTCVczxFTlfdMd502PjcMiOOzNG39u4N8F68-9m8D_NNtwaJTXk6Fng_YNADL9mTJPpfXLY5SQY5mQY1xqA6brO6Nl_6C7MccjxTDanlQApIVNsoVwMKUVsD2YomLuuzNKVyV2Z-67MPh-9exrjcPK3nAzwBUh5NWwxPv79jOwf5CmgzQ</recordid><startdate>20201201</startdate><enddate>20201201</enddate><creator>Miyagi, Chihiro</creator><creator>Ochiai, Yoshie</creator><creator>Ando, Yusuke</creator><creator>Hisahara, Manabu</creator><creator>Baba, Hironori</creator><creator>Takigawa, Tomoya</creator><creator>Tokunaga, Shigehiko</creator><general>Springer Singapore</general><general>Springer Nature B.V</general><scope>C6C</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>3V.</scope><scope>7X7</scope><scope>7XB</scope><scope>8FI</scope><scope>8FJ</scope><scope>8FK</scope><scope>ABUWG</scope><scope>AFKRA</scope><scope>BENPR</scope><scope>CCPQU</scope><scope>FYUFA</scope><scope>GHDGH</scope><scope>K9.</scope><scope>M0S</scope><scope>PQEST</scope><scope>PQQKQ</scope><scope>PQUKI</scope><scope>7X8</scope><scope>5PM</scope></search><sort><creationdate>20201201</creationdate><title>A case of cardiac strangulation following epicardial pacemaker implantation</title><author>Miyagi, Chihiro ; Ochiai, Yoshie ; Ando, Yusuke ; Hisahara, Manabu ; Baba, Hironori ; Takigawa, Tomoya ; Tokunaga, Shigehiko</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c564t-4631ed2ecf0083e89b763ced3c7b8e07be164f765c4291627e4967f46df4c52c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2020</creationdate><topic>Age</topic><topic>Cardiac arrhythmia</topic><topic>Cardiac Pacing, Artificial</topic><topic>Cardiac Surgery</topic><topic>Cardiology</topic><topic>Case Report</topic><topic>Catheters</topic><topic>Child</topic><topic>Coronary vessels</topic><topic>Heart</topic><topic>Heart Atria</topic><topic>Heart Block - therapy</topic><topic>Heart Ventricles</topic><topic>Humans</topic><topic>Male</topic><topic>Medical imaging</topic><topic>Medicine</topic><topic>Medicine & Public Health</topic><topic>Pacemaker, Artificial - adverse effects</topic><topic>Pacemakers</topic><topic>Pericardium</topic><topic>Pulmonary arteries</topic><topic>Strangulation</topic><topic>Surgical Oncology</topic><topic>Thoracic Surgery</topic><topic>Tomography</topic><topic>Veins & arteries</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Miyagi, Chihiro</creatorcontrib><creatorcontrib>Ochiai, Yoshie</creatorcontrib><creatorcontrib>Ando, Yusuke</creatorcontrib><creatorcontrib>Hisahara, Manabu</creatorcontrib><creatorcontrib>Baba, Hironori</creatorcontrib><creatorcontrib>Takigawa, Tomoya</creatorcontrib><creatorcontrib>Tokunaga, Shigehiko</creatorcontrib><collection>Springer Nature OA Free Journals</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Central (Corporate)</collection><collection>Health & Medical Collection</collection><collection>ProQuest Central (purchase pre-March 2016)</collection><collection>Hospital Premium Collection</collection><collection>Hospital Premium Collection (Alumni Edition)</collection><collection>ProQuest Central (Alumni) (purchase pre-March 2016)</collection><collection>ProQuest Central (Alumni Edition)</collection><collection>ProQuest Central UK/Ireland</collection><collection>ProQuest Central</collection><collection>ProQuest One Community College</collection><collection>Health Research Premium Collection</collection><collection>Health Research Premium Collection (Alumni)</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Health & Medical Collection (Alumni Edition)</collection><collection>ProQuest One Academic Eastern Edition (DO NOT USE)</collection><collection>ProQuest One Academic</collection><collection>ProQuest One Academic UKI Edition</collection><collection>MEDLINE - Academic</collection><collection>PubMed Central (Full Participant titles)</collection><jtitle>General thoracic and cardiovascular surgery</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Miyagi, Chihiro</au><au>Ochiai, Yoshie</au><au>Ando, Yusuke</au><au>Hisahara, Manabu</au><au>Baba, Hironori</au><au>Takigawa, Tomoya</au><au>Tokunaga, Shigehiko</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A case of cardiac strangulation following epicardial pacemaker implantation</atitle><jtitle>General thoracic and cardiovascular surgery</jtitle><stitle>Gen Thorac Cardiovasc Surg</stitle><addtitle>Gen Thorac Cardiovasc Surg</addtitle><date>2020-12-01</date><risdate>2020</risdate><volume>68</volume><issue>12</issue><spage>1499</spage><epage>1502</epage><pages>1499-1502</pages><issn>1863-6705</issn><eissn>1863-6713</eissn><abstract>An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further studies including computed tomography and cardiac catheterization revealed that the pacemaker lead had completely encircled the cardiac silhouette and was in a state of “cardiac strangulation”. We removed the previous pacing leads and generator and implanted a new epicardial dual-chamber pacing system in the right atrium and right ventricle. Additionally, an expanded polytetrafluoroethylene sheet was placed between the new leads and the heart to prevent recurrence of cardiac strangulation.</abstract><cop>Singapore</cop><pub>Springer Singapore</pub><pmid>32270437</pmid><doi>10.1007/s11748-020-01337-y</doi><tpages>4</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1863-6705 |
ispartof | General thoracic and cardiovascular surgery, 2020-12, Vol.68 (12), p.1499-1502 |
issn | 1863-6705 1863-6713 |
language | eng |
recordid | cdi_pubmedcentral_primary_oai_pubmedcentral_nih_gov_7680322 |
source | MEDLINE; SpringerLink Journals - AutoHoldings; ProQuest Central |
subjects | Age Cardiac arrhythmia Cardiac Pacing, Artificial Cardiac Surgery Cardiology Case Report Catheters Child Coronary vessels Heart Heart Atria Heart Block - therapy Heart Ventricles Humans Male Medical imaging Medicine Medicine & Public Health Pacemaker, Artificial - adverse effects Pacemakers Pericardium Pulmonary arteries Strangulation Surgical Oncology Thoracic Surgery Tomography Veins & arteries |
title | A case of cardiac strangulation following epicardial pacemaker implantation |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-28T21%3A58%3A35IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20case%20of%20cardiac%20strangulation%20following%20epicardial%20pacemaker%20implantation&rft.jtitle=General%20thoracic%20and%20cardiovascular%20surgery&rft.au=Miyagi,%20Chihiro&rft.date=2020-12-01&rft.volume=68&rft.issue=12&rft.spage=1499&rft.epage=1502&rft.pages=1499-1502&rft.issn=1863-6705&rft.eissn=1863-6713&rft_id=info:doi/10.1007/s11748-020-01337-y&rft_dat=%3Cproquest_pubme%3E2918740660%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2918740660&rft_id=info:pmid/32270437&rfr_iscdi=true |