A case of cardiac strangulation following epicardial pacemaker implantation
An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further st...
Gespeichert in:
Veröffentlicht in: | General thoracic and cardiovascular surgery 2020-12, Vol.68 (12), p.1499-1502 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | An 8-year-old boy had undergone permanent epicardial pacemaker implantation with a Y-shaped bipolar ventricular lead on day 6 after birth for treatment of congenital complete atrioventricular block. He was found to have pulmonary stenosis and mitral stenosis by follow-up echocardiography. Further studies including computed tomography and cardiac catheterization revealed that the pacemaker lead had completely encircled the cardiac silhouette and was in a state of “cardiac strangulation”. We removed the previous pacing leads and generator and implanted a new epicardial dual-chamber pacing system in the right atrium and right ventricle. Additionally, an expanded polytetrafluoroethylene sheet was placed between the new leads and the heart to prevent recurrence of cardiac strangulation. |
---|---|
ISSN: | 1863-6705 1863-6713 |
DOI: | 10.1007/s11748-020-01337-y |