Purpose Mixed reality (MR) is being evaluated as a visual tool for surgical navigation. Current literature presents unclear results on intraoperative accuracy using the Microsoft HoloLens 1 ® . This study aims to assess the impact of the surgeon’s sightline in an inside-out marker-based MR navigatio...
Gespeichert in:
Veröffentlicht in: | International journal for computer assisted radiology and surgery 2020-12, Vol.15 (12), p.2027-2039 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Mixed reality (MR) is being evaluated as a visual tool for surgical navigation. Current literature presents unclear results on intraoperative accuracy using the Microsoft HoloLens 1
®
. This study aims to assess the impact of the
surgeon’s sightline
in an inside-out marker-based MR navigation system for open surgery.
Methods
Surgeons at Akershus University Hospital tested this system. A custom-made phantom was used, containing 18 wire target crosses within its inner walls. A CT scan was obtained in order to segment all wire targets into a single 3D-model (hologram). An in-house software application (CTrue), developed for the Microsoft HoloLens 1, uploaded 3D-models and automatically registered the 3D-model with the phantom. Based on the surgeon’s sightline while registering and targeting (free sightline /F/or a strictly perpendicular sightline /P/), 4 scenarios were developed (FF-PF-FP-PP). Target error distance (TED) was obtained in three different working axes-(
XYZ
).
Results
Six surgeons (5 males, age 29–62) were enrolled. A total of 864 measurements were collected in 4 scenarios, twice. Scenario PP showed the smallest TED in
XYZ
-axes mean = 2.98 mm ± SD 1.33; 2.28 mm ± SD 1.45; 2.78 mm ± SD 1.91, respectively. Scenario FF showed the largest TED in
XYZ
-axes with mean = 10.03 mm ± SD 3.19; 6.36 mm ± SD 3.36; 16.11 mm ± SD 8.91, respectively. Multiple comparison tests, grouped in scenarios and axes, showed that the majority of scenario comparisons had significantly different TED values (