Conditional safety margins for less conservative peak local SAR assessment: A probabilistic approach
Purpose The introduction of a linear safety factor to address peak local specific absorption rate (pSAR10g) uncertainties (eg, intersubject variation, modeling inaccuracies) bears one considerable drawback: It often results in over‐conservative scanning constraints. We present a more efficient appro...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance in medicine 2020-12, Vol.84 (6), p.3379-3395 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
The introduction of a linear safety factor to address peak local specific absorption rate (pSAR10g) uncertainties (eg, intersubject variation, modeling inaccuracies) bears one considerable drawback: It often results in over‐conservative scanning constraints. We present a more efficient approach to define a variable safety margin based on the conditional probability density function of the effectively obtained pSAR10g value, given the estimated pSAR10g value.
Methods
The conditional probability density function can be estimated from previously simulated data. A representative set of true and estimated pSAR10g samples was generated by means of our database of 23 subject‐specific models with an 8‐fractionated dipole array for prostate imaging at 7 T. The conditional probability density function was calculated for each possible estimated pSAR10g value and used to determine the corresponding safety margin with an arbitrary low probability of underestimation. This approach was applied to five state‐of‐the‐art local SAR estimation methods, namely: (1) using just the generic body model “Duke”; (2) using our model library to assess the maximum pSAR10g value over all models; (3) using the most representative “local SAR model”; (4) using the five most representative local SAR models; and (5) using a recently developed deep learning–based method.
Results
Compared with the more conventional safety factor, the conditional safety‐margin approach results in lower (up to 30%) mean overestimation for all investigated local SAR estimation methods.
Conclusion
The proposed probabilistic approach for pSAR10g correction allows more accurate local SAR assessment with much lower overestimation, while a predefined level of underestimation is accepted (eg, 0.1%). |
---|---|
ISSN: | 0740-3194 1522-2594 |
DOI: | 10.1002/mrm.28335 |