QuantiFERON-TB Gold Plus with Chemiluminescence Immunoassay: Do We Need a Higher Cutoff?
QuantiFERON-TB Gold Plus (QFT-Plus) is the most widely used interferon gamma release assay (IGRA) for the diagnosis of latent tuberculosis infection (LTBI). The aim of this study was to compare QFT-Plus results by enzyme-linked immunosorbent assay (ELISA) on the SkyLab system with those obtained wit...
Gespeichert in:
Veröffentlicht in: | Journal of clinical microbiology 2020-09, Vol.58 (10) |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | QuantiFERON-TB Gold Plus (QFT-Plus) is the most widely used interferon gamma release assay (IGRA) for the diagnosis of latent tuberculosis infection (LTBI). The aim of this study was to compare QFT-Plus results by enzyme-linked immunosorbent assay (ELISA) on the SkyLab system with those obtained with chemiluminescence immunoassay (CLIA) on the Liaison XL analyzer. Agreement between the two assays was evaluated on 419 QFT-Plus blood samples and was found to be substantial (75.4%); higher agreement was found for positive (95.4%) and negative (80.4%) results, while most discordances were due to ELISA-indeterminate/CLIA-determinate results. According to Italian Clinical Microbiologist Association recommendations, in samples (
= 79) with a borderline result in ELISA (0.20 to 0.70 IU/ml), CLIA median values statistically increased (from 0.29 to 0.59 IU/ml for TB1 and from 0.32 to 0.60 IU/ml for TB2) but remained in the borderline range. Linear regression analysis indicated a substantial correlation between ELISA and CLIA for antigen tubes TB1 (Pearson's
= 0.8666) and TB2 (Pearson's
= 0.8728), but CLIA produced higher values than ELISA. Receiver operating characteristic (ROC) analysis showed that the optimal cutoff value in CLIA was 0.45 IU/ml for TB1 and 0.46 IU/ml for TB2. In conclusion, automated QFT-Plus with CLIA is comparable to QFT-Plus performed by ELISA. Within the linearity range of the test, CLIA detects higher quantitative values than ELISA, resulting in a higher number of determinate results and the conversion of samples that were close to the cutoff into positive borderline results. A higher cutoff for QFT-CLIA needs to be defined based on clinical diagnostic criteria. |
---|---|
ISSN: | 0095-1137 1098-660X |
DOI: | 10.1128/JCM.00780-20 |