Effect of differences in pixel size on image characteristics of digital intraoral radiographic systems: a physical and visual evaluation

To quantify and validate the effect of pixel size on a digital intraoral radiographic system according to International Electrotechnical Commission standards through physical and visual evaluations. The digital intraoral radiographic system used was the photostimulable phosphor imaging plate and sca...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Dento-maxillo-facial radiology 2020-09, Vol.49 (6), p.20190378-20190378
Hauptverfasser: Kuramoto, Taku, Takarabe, Shinya, Okamura, Kazutoshi, Shiotsuki, Kenshi, Shibayama, Yusuke, Tsuru, Hiroki, Akamine, Hiroshi, Tatsumi, Masato, Kato, Toyoyuki, Morishita, Junji, Yoshiura, Kazunori
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:To quantify and validate the effect of pixel size on a digital intraoral radiographic system according to International Electrotechnical Commission standards through physical and visual evaluations. The digital intraoral radiographic system used was the photostimulable phosphor imaging plate and scanner system. The system had three image capture modes: high-speed (HS), high-resolution (HR), and super high-resolution (SHR) with different pixels. The physical characteristics of the system were evaluated using presampled modulation transfer function (MTF) and the normalized noise power spectrum (NNPS). An aluminum (Al) step phantom with different depths of holes was used to acquire images under various exposure conditions. The average number of perceptible holes from all steps was plotted against each exposure dose. The results were compared to analyze the effects of pixel size on image quality of intraoral radiographs. The MTF was slightly higher with SHR than with HR and HS. The NNPS with SHR showed about a 40% decrease in magnitude compared to HS. The total number of perceptible holes in the Al step phantom was higher with SHR than with HS and HR in all exposure conditions. The MTF and NNPS obtained with different pixel size could be quantified by physical evaluation, and the differences were visually validated with Al step phantom. The SHR mode has the potential to decrease the radiation dose without compromising the image quality.
ISSN:0250-832X
1476-542X
DOI:10.1259/DMFR.20190378