Use of the variplex™ SARS-CoV-2 RT-LAMP as a rapid molecular assay to complement RT-PCR for COVID-19 diagnosis
•Rapid detection of SARS-CoV-2 by variplex™ RT-LAMP from respiratory samples.•Homogenization of samples using SL solution for testing without RNA elution.•Combination of RT-LAMP and RT-PCR increases diagnostic accuracy. Molecular assays based on reverse transcription-loop-mediated isothermal amplifi...
Gespeichert in:
Veröffentlicht in: | Journal of clinical virology 2020-11, Vol.132, p.104616-104616, Article 104616 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | •Rapid detection of SARS-CoV-2 by variplex™ RT-LAMP from respiratory samples.•Homogenization of samples using SL solution for testing without RNA elution.•Combination of RT-LAMP and RT-PCR increases diagnostic accuracy.
Molecular assays based on reverse transcription-loop-mediated isothermal amplification (RT-LAMP) may be useful for rapid diagnosis of the severe acute respiratory syndrome Coronavirus-2 (SARS-CoV-2) because of the easy performance and the option to bypass RNA extraction.
This study was designed to evaluate the clinical performance of the CE-labeled variplexTM real time SARS-CoV-2 RT-LAMP assay in comparison to commercial RT-PCRs.
RNA extracted from pharyngeal swabs was tested by variplex™ RT-LAMP and Corman’s LightMix™ E gene RT-PCR as reference. Samples of respiratory secretions from Coronavirus infection disease (COVID-19) and negative control patients were analyzed by variplex™ without RNA extraction and tested in parallel with the Allplex™ and VIASURE BD MAX RT-PCRs.
Using isolated RNA variplex™ RT-LAMP showed a sensitivity of 75 % compared to LightMix E gene RT-PCR but contrary to the latter it produced no false-positive results. For the evaluation of samples from respiratory secretions concordance analysis showed only a moderate agreement between the variplex™ RT-LAMP conducted on unprocessed samples and Allplex™ and VIASURE RT-PCRs (Cohen’s κ ranging from 0.52−0.56). Using the approach to define a sample as true-positive when at least two assays gave a positive result the clinical sensitivities were as follows: 76.3 % for variplex™, 84.2 % for Allplex™ and 68.4 % for VIASURE. However, when results of RT-PCR and RT-LAMP were combined diagnostic sensitivity was increased to 92–100 %.
The variplex RT-LAMP may serve as a rapid test to be combined with a RT-PCR assay to increase the diagnostic accuracy in patients with suspected COVID-19 infection. |
---|---|