Comparison of angle-to-angle distance using three devices in normal eyes
Purpose To compare the angle-to-angle (ATA) distance, and the repeatability and the reproducibility of these measurements among the three devices. Methods We performed the ATA measurements in 26 healthy subjects using anterior segment optical coherence tomography (CASIA2, Tomey, Japan), the Scheimpf...
Gespeichert in:
Veröffentlicht in: | Eye (London) 2020-06, Vol.34 (6), p.1116-1120 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To compare the angle-to-angle (ATA) distance, and the repeatability and the reproducibility of these measurements among the three devices.
Methods
We performed the ATA measurements in 26 healthy subjects using anterior segment optical coherence tomography (CASIA2, Tomey, Japan), the Scheimpflug camera (Pentacam HR, Oculus, Wetzlar, Germany), and the combined Placido-ring corneal topography and Scheimpflug camera (TMS-5, Tomey). We also compared the repeatability and the reproducibility of the ATA measurements among these three devices.
Results
The ATA in the CASIA2 group was significantly larger than that in the Pentacam group (Bonferroni test,
p
= 0.002), or that in the TMS-5 group (
p
|
---|---|
ISSN: | 0950-222X 1476-5454 |
DOI: | 10.1038/s41433-019-0653-2 |