Production and clinical evaluation of breast lesion skin markers for automated three-dimensional ultrasonography of the breast: a pilot study
Objectives Automated ultrasound of the breast has the advantage to have the whole breast scanned by technicians. Consequently, feedback to the radiologist about concurrent focal abnormalities (e.g., palpable lesions) is lost. To enable marking of patient- or physician-reported focal abnormalities, w...
Gespeichert in:
Veröffentlicht in: | European radiology 2020-06, Vol.30 (6), p.3356-3362 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Objectives
Automated ultrasound of the breast has the advantage to have the whole breast scanned by technicians. Consequently, feedback to the radiologist about concurrent focal abnormalities (e.g., palpable lesions) is lost. To enable marking of patient- or physician-reported focal abnormalities, we aimed to develop skin markers that can be used without disturbing the interpretability of the image.
Methods
Disk-shaped markers were casted out of silicone. In this IRB-approved prospective study, 16 patients were included with a mean age of 57 (39–85). In all patients, the same volume was imaged twice using an automated breast ultrasound system, once with and once without a marker in place. Nine radiologists from two medical centers filled scoring forms regarding image quality, image interpretation, and confidence in providing a diagnosis based on the images.
Results
Marker adhesion was sufficient for automated scanning. Observer scores showed a significant shift in scores from excellent to good regarding diagnostic yield/image quality (
χ
2
, 15.99,
p
|
---|---|
ISSN: | 0938-7994 1432-1084 |
DOI: | 10.1007/s00330-020-06695-y |