Antimicrobial Silver Multilayer Coating for Prevention of Bacterial Colonization of Orthopedic Implants

Due to increasing rates of periprosthetic joint infections (PJI), new approaches are needed to minimize the infection risk. The first goal of this study was to modify a well-established infection model to test surface-active antimicrobial systems. The second goal was to evaluate the antimicrobial ac...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Materials 2020-03, Vol.13 (6), p.1415
Hauptverfasser: Fabritius, Martin, Al-Munajjed, Amir Andreas, Freytag, Christiane, Jülke, Henriette, Zehe, Markus, Lemarchand, Thomas, Arts, Jacobus J, Schumann, Detlef, Alt, Volker, Sternberg, Katrin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Due to increasing rates of periprosthetic joint infections (PJI), new approaches are needed to minimize the infection risk. The first goal of this study was to modify a well-established infection model to test surface-active antimicrobial systems. The second goal was to evaluate the antimicrobial activity of a silver multilayer (SML) coating. In vitro tests with SML items showed a >4 Log reduction in a proliferation assay and a 2.2 Log reduction in an agar immersion test (7 d). In the in vivo model blank and SML coated K-wires were seeded with ~2 × 10 CFU of a methicillin-sensitive Staphylococcus epidermidis (MSSE) and inserted into the intramedullary tibial canal of rabbits. After 7 days, the animals were sacrificed and a clinical, microbiological and histological analysis was performed. Microbiology showed a 1.6 Log pathogen reduction on the surface of SML items ( = 0.022) and in loosely attached tissue ( = 0.012). In the SML group 7 of 12 SML items were completely free of pathogens (cure rate = 58%, = 0.002), while only 1 of 12 blank items were free of pathogens (cure rate = 8%, = 0.110). No silver was detected in the blood or urine of the SML treated animals and only scarcely in the liver or adjacent lymph nodes. In summary, an in vivo infection model to test implants with bacterial pre-incubation was established and the antimicrobial activity of the SML coating was successfully proven.
ISSN:1996-1944
1996-1944
DOI:10.3390/ma13061415