Implementing a negative-pressure isolation ward for a surge in airborne infectious patients

Highlights • A 30-bed negative-pressure isolation ward was established on a functioning hospital. • The pressure relative to the main hospital was −29 Pa by adjusting the ventilation. • No occurrences of pressure reversal occurred at ward entrance. • Pressures on the ward changed to slightly positiv...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of infection control 2017-06, Vol.45 (6), p.652-659
Hauptverfasser: Miller, Shelly L., PhD, Clements, Nicholas, PhD, Elliott, Steven A, Subhash, Shobha S., MS, Eagan, Aaron, MS, Radonovich, Lewis J., MD
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Highlights • A 30-bed negative-pressure isolation ward was established on a functioning hospital. • The pressure relative to the main hospital was −29 Pa by adjusting the ventilation. • No occurrences of pressure reversal occurred at ward entrance. • Pressures on the ward changed to slightly positive. • Health care personnel should wear personal protective equipment on the ward.
ISSN:0196-6553
1527-3296
DOI:10.1016/j.ajic.2017.01.029