Rapid Antigen Tests for Diagnosis of Pandemic (Swine) Influenza A/H1N1
We found that the sensitivities of 3 rapid influenza antigen tests for pandemic influenza A/H1N1 virus were low to moderate: BD Directigen EZ Flu A+B test (Becton Dickinson), 46.7%; BinaxNOW Influenza A&B (Inverness Medical), 38.3%; and QuickVue Influenza A+B Test (Quidel), 53.3%. A patient with...
Gespeichert in:
Veröffentlicht in: | Clinical infectious diseases 2009-10, Vol.49 (7), p.1090-1093 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We found that the sensitivities of 3 rapid influenza antigen tests for pandemic influenza A/H1N1 virus were low to moderate: BD Directigen EZ Flu A+B test (Becton Dickinson), 46.7%; BinaxNOW Influenza A&B (Inverness Medical), 38.3%; and QuickVue Influenza A+B Test (Quidel), 53.3%. A patient with influenza-like illness who has a negative rapid antigen test result should undergo further testing using reverse-transcription polymerase chain reaction. |
---|---|
ISSN: | 1058-4838 1537-6591 |
DOI: | 10.1086/644743 |