Rapid Antigen Tests for Diagnosis of Pandemic (Swine) Influenza A/H1N1

We found that the sensitivities of 3 rapid influenza antigen tests for pandemic influenza A/H1N1 virus were low to moderate: BD Directigen EZ Flu A+B test (Becton Dickinson), 46.7%; BinaxNOW Influenza A&B (Inverness Medical), 38.3%; and QuickVue Influenza A+B Test (Quidel), 53.3%. A patient with...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical infectious diseases 2009-10, Vol.49 (7), p.1090-1093
Hauptverfasser: Vasoo, Shawn, Stevens, Jane, Singh, Kamaljit
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:We found that the sensitivities of 3 rapid influenza antigen tests for pandemic influenza A/H1N1 virus were low to moderate: BD Directigen EZ Flu A+B test (Becton Dickinson), 46.7%; BinaxNOW Influenza A&B (Inverness Medical), 38.3%; and QuickVue Influenza A+B Test (Quidel), 53.3%. A patient with influenza-like illness who has a negative rapid antigen test result should undergo further testing using reverse-transcription polymerase chain reaction.
ISSN:1058-4838
1537-6591
DOI:10.1086/644743