Evaluation of Mechanical Performances of Stents with 38 mm Length in Long Lesion

Objective. Single stent with 38 mm length has emerged as a potential solution for long lesion treatment using PCI. However, long stents need to come over a longer lesion length combined with a higher incidence of tortuous calcification, requiring a stent to provide superior transport and compliance...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:BioMed research international 2020, Vol.2020 (2020), p.1-6
Hauptverfasser: Zhang, Jinying, Shen, Deliang, Zhang, Zenglei, Cao, Chang, Zhang, Jianchao, Guo, Jiacheng, Yue, Xiaoting, Tang, Junnan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Objective. Single stent with 38 mm length has emerged as a potential solution for long lesion treatment using PCI. However, long stents need to come over a longer lesion length combined with a higher incidence of tortuous calcification, requiring a stent to provide superior transport and compliance and reduce elastic retraction. Here, we evaluated the mechanical performances of the existing four types of drug eluting stents with 38 mm length, which could provide guidance for clinicians to choose the proper stents for the patients. Methods. The stents with 38 mm length from XIENCE Xpedition (Abbott, US), SYNERGY (Boston Scientific, US), FIREHAWK (Microport, China), and HELIOS (HELIOS, China) were collected. Mechanical parameters of stents including crossing ability, compliance, elastic recoil, and longitudinal strength were performed. Results. The resistance force of stents from XIENCE Xpedition was smaller than FIREHAWK (p
ISSN:2314-6133
2314-6141
DOI:10.1155/2020/2594161